User guide
The following table lists the accessible functions at any key-pad lock-out
level (L0or L1) both when the pump is in OFF and ON condition.
SECTION
7
35
TABLE OF KEY-PAD FUNCTIONS
MAN 01/06/USA/p CRONO FIVE 11.10
FUNCTIONS AND SELECTIONS
Switching ON the pump
Switching OFF the pump
Prime
Bolus dose
Displaying delivery time
Displaying volume delivered as flow rate
Displaying volume delivered as bolus dose
Displaying number of bolus dose delivered
Clearing the number of bolus doses delivered
Displaying volume delivered as clinician bolus
Displaying total volume delivered
Displaying programmed parameters
Flow rate setting
Bolus dose setting
Interval between bolus doses setting
Number of bolus dose in one hour setting
Partial volume setting
Displaying the number of complete infusions
Clearing the number of complete infusions
Lock/Unlock the keyboard
Clinician bolus
OFF
ON
L0 L1 L0 L1
YESYES NO NO
YESYES NO NO
YESYES NO NO
YESYES NO NO
NO NO YES YES
NO NO YES YES
NO NO YES YES
NO NO YES YES
NO NO YES YES
NO NO YES YES
NO NO YES YES
NO NO YES YES
NO NO YES YES
YES NO NO NO
YES NO NO NO
YES NO NO NO
YES NO NO NO
YES NO NO NO
YES NO NO NO
YES NO NO NO
NO NO YES*YES*
*
Available only at the beginning of a new infusion or a new partial infusion.