User guide

SECTION 8
36
PROGRAMMING
MAN 01/06/USA/p CRONO FIVE 11.10
For programming, the pump must be:
In OFF condition.
With key-pad lock-out in L0.
Depress the button for about 1 second to access the pump program -
ming sequence. The first programming option displayed on the screen is
flow rate. The flow rate can be changed with the button or the but-
ton as long as the presently programmed flow rate is flashing.
The second depressing of the button will display the second program-
ming option bolus dose volume. The bolus dose volume can be changed
with the button or the button as long as the presently program -
med bolus dose volume is flashing.
The third depressing of the button will display the third programming
option minimum time interval between bolus doses. The minimum time
interval between bolus doses can be changed with the button or the
button as long as the presently programmed interval time is flashing.
The fourth depressing of the button will display the fourth program -
ming option number of bolus doses per hour . The number of bolus
doses per hour can be changed with the button or the button as
long as the presently programmed number of bolus doses per hour is flas-
hing.
The fifth depressing of the button will display the fifth programming
option partial volume. The partial volume can be changed with the
button or the button as long as the presently programmed partial volu-
me is flashing. The partial volume can only be programmed when the pus-
her is in the start position.
NOTE
If the key pad lock-out is programmed to L1, depressing the button will
display L1 (key pad lock-out: programming is not allowed).