PROGRAMMING OF FLOW RATE
MAN 01/06/USA/p CRONO FIVE 11.10
The flow rate can be programmed from 5
mcl/h to 5000 mcl/h as follows:
From 5 mcl/h to 1000 mcl/h with increments
of 5 mcl/h.
From 1000 mcl/h to 5000 mcl/h with incre-
ments of 10 mcl/h.
Programming procedure:
1. When the button is depressed for
about 1 second with the pump in OFF
condition, the presently programmed
flow rate will be displayed and flashing
2 . By depressing the button the value will
increase; by depressing the button
the value will decrease. Any change in
value is followed by an acoustic signal.
3. If no button is depressed within 8 seconds
in the programming phase, the value stops
flashing and OFF appears on the screen.
4. Depressing the button before the OFF
message appears (the flow rate value is
still flashing) moves the programming
sequence to the next option: bolus dose.
NOTE
You can accelerate (scrolling) the display of
new values by depressing the or the
button continuously (in increments of 20 or
40 mcl/h depending on the range).
Programmed flow rates are automatically
stored.