PROGRAMMING OF BOLUS DOSE
MAN 01/06/USA/p CRONO FIVE 11.10
10
20
10
20
10
20
The bolus dose can be programmed from 0.01
ml to 2.00 ml with increments of 0.01 ml.
The bolus dose is accessed by the second
depressing of the button in the pump pro-
gramming sequence (see page 37).
Programming procedure:
1. When the button is depressed for a
second time, a "d" is displayed and the
presently programmed bolus dose is flas-
hing.
2. By depressing the button the value will
increase; by depressing the button the
value will decrease. Any change in value
is followed by an acoustic signal.
3. If no button is depressed within 8 seconds
in the programming phase, the value stops
flashing and OFF appears on the screen.
4. Depressing the button before the OFF
message appears (the bolus dose is still
flashing) moves the programming sequen-
ce to the next option: minimum time
interval between bolus doses.
NOTE
You can accelerate (scrolling) the display of
new values by depressing the or the
button continuously (in increments of 0,04
ml).
Programmed bolus doses are automatically
stored.
By programming d 0,00, the bolus dose is
eliminated.