User guide
SECTION
8
40
MINIMUM TIME INTERVAL BETWEEN BOLUS DOSES
MAN 01/06/USA/p CRONO FIVE 11.10
10
20
10
20
The minimum time interval between bolus
doses can be programmed from 5 minutes to
24 hours as follows:
• From 5 minutes to 1 hour with increments of
5 minutes.
• From 1 hour to 24 hours with increments of
15 minutes.
This function limits the patient to access the
programmed bolus dose before a certain time
has elapsed from the latest administered
bolus dose.
The minimum time interval between bolus
doses is accessed by the third depressing of
the button in the pump programming
sequence (see page 37).
Programming procedure:
1. When the button is depressed for a
third time, a "t" is displayed and the pre-
sently programmed interval time is
flashing.
2. By depressing the button the value
will increase; by depressing the button
the value will decrease. Any change in
value is followed by an acoustic signal.
3. If no button is depressed within 8 seconds
in the programming phase, the value stops
flashing and OFF appears on the screen.
4. Depressing the button before the OFF
message appears (the interval time is still
flashing) moves the programming
sequence to the next option: bolus doses
per hour.