User guide

SECTION
8
44
PROGRAMMING OF PARTIAL VOLUME
MAN 01/06/USA/p CRONO FIVE 11.10
10
20
10
20
10
20
Partial volume is used when less than 20 ml is
drawn up in the syringe for a specific treatment
or for a specific patient, for example children
who needs less than the adult dose.
Partial volume can be programmed between 1
ml and 20 ml with increments of 1 ml.
Partial volume is accessed by the fifth depres-
sing of the button in the pump program -
ming sequence (see page 37).
Please observe that partial volume is only
available at the beginning of a new or new par-
tial infusion.
Programming procedure:
1. When the button is depressed for a
fifth time, the presently programmed volu-
me in ml is flashing:
2. By depressing the button the value will
increase; by depressing the button the
value will decrease. Any change in value is
followed by an acoustic signal.
3. If no button is depressed within 8 seconds
in the programming phase, the selected
volume in ml stops flashing and P,cc
appears on the screen.
4. The pusher will automatically position itself
for the programmed partial volume. During
the movement of the pusher , there will be
an acoustic signal and the delivery time
corresponding to the selected partial volu-
me will be displayed on the screen.
5. When the pusher reaches its programmed
position, the display will show OFF.