User guide

SECTION 9
48
MAN 01/06/USA/p CRONO FIVE 11.10
b. To skip priming
Depress the button; the pump will start the
infusion and the display will show the remai -
ning infusion time.
c. To start priming
Depress the button: the pump will deliver
0,5 ml and the display will show a P followed
by the volume in ml that has been primed.
After the first priming of 0.5 ml, the display will
show Pr again and the procedure can be
repeated two more times up to 1,5 ml has
been primed.
There is no time limit to finish the priming pro-
cedure.
NOTE
If no button is depressed for 10 seconds in
Pr, the pump will switch to OFF.
The priming volume is not included in the
administration volume.
The priming can be interrupted at any time
by depressing the button and the but-
ton simultaneously. The display will show
Pr and alternative a, b or c can be selec-
ted again.
10
20
10
20
10
20
10
20