BOLUS DOSE LIMITATIONS
MAN 01/06/USA/p CRONO FIVE 11.10
• Time interval between bolus doses
(t limitation)
• Number of bolus doses per hours
(n limitation)
Bolus doses cannot be delivered before the
time (t) has expired or when the number of
programmed boluses per hour (n) has
already been administered.
The pump will always choose the program -
med (t) or (n) that administer the lowest
amount of bolus doses in ml.
If a bolus dose is requested during limitation
time (t) or after the maximum number of bolus
doses (n) has been reached, the display will
show error message  Er,t  and Er,n respecti-
vely.