User guide

CAUTIONS
Disconnect the infusion
set from the patient or
put a clamp on the infu-
sion set before starting
the reversal of the pus-
her to avoid that medi-
cation is aspirated from
the infusion set.
Do not remove the syrin-
ge before the pusher
has reached the starting
position for a new infu-
sion.
SECTION 9
56
MAN 01/06/USA/p CRONO FIVE 11.10
the and the buttons simultaneously;
OFF will be displayed; End and OFF will
alternate on the screen.
By depressing the button, the pump will
resume the reversal of the pusher.
If pump reversal is requested when the
pusher already is in the starting position for
an infusion, the pump will emit an acoustic
signal and Err will be displayed.
The reversal of the pusher at the end of
infusion can be interrupted by depressing
the and the buttons simultaneously;
End and OFF will alternate on the screen.
By depressing the button, the pump will
resume the reversal of the pusher.