User guide

SECTION 10
60
DISPLAYING DELIVERED VOLUMES
MAN 01/06/USA/p CRONO FIVE 11.10
This display procedure will show the delivered
volume in ml since the start of the infusion for
the different delivery options.
The pump has to be running (ON) and the
screen will either show the remaining delivery
time or F0,00.
Display procedure:
1. Depress the button for about 1 second
and the programmed flow rate will be
displayed.
2. If no button is depressed for 5 seconds, the
pump will revert to remaining delivery time
or F0,00.
3. If the button is depressed again within
the 5 second time interval, the next value
will be displayed as follows: Volume delive-
red as bolus doses, Number of bolus
doses delivered, Volume delivered as
clinician boluses, Total volume delive-
red.
NOTE
The lock level of the key-pad lock out does
not interfere with programmed values.
The values of the volumes delivered are
constantly updated during the infusion.
All values related to delivered volume and
number of bolus doses are automatically
reset to zero at the beginning of a new or
partial new infusion.
The chart beside shows the sequence of the
screens related to the volumes delivered by
the pump during an infusion.