User guide

SECTION 10
63
DISPLAYING THE NUMBER OF COMPLETE INFUSIONS
MAN 01/06/USA/p CRONO FIVE 11.10
The pump has to be in OFF to be able to dis -
play the number of complete infusions perfor-
med by the pump.
Depress the button for about 5 seconds
and the number of complete infusions will be
displayed.
Clearing the number of complete infusions
performed by the pump.
Clearing procedure:
1. The pump has to be in OFF and lock level
L0.
2. Depress the button for about 5 seconds
and the number of complete infusions will
be displayed.
3. Without releasing the button, depress
the button as well and the number of
complete infusions will start flashing on the
screen.
4. Depress the button for about 1 second
and thereafter the button for about 1
second (in sequence) and the button
within 7 seconds. The pump will emit a tic -
king sound followed by a longer signal.
Then the number of complete infusions is
cleared.