User guide

WARNINGS
Do not use rechargeable
batteries.
The use of batteries of a
different type could
result in device malfunc-
tions.
Battery life will vary
depending on the battery
age, temperature and
storage conditions.
Make sure that a new
battery is kept available
for emergency replace-
ment.
Do not store the pump for
prolonged periods (1-2
months) with the battery
installed.
Make sure that the bat-
tery door is properly clo-
sed.
Dispose of used batteries
in an environmentally
safe manner and accor-
dingly to any regulations
which may apply.
SECTION 11
67
MAN 01/06/USA/p CRONO FIVE 11.10
If it is dif ficult to remove the battery using the
ribbon strap, do not use any other object to
remove the battery but try the following proce-
dure:
Hold the pump and the battery compart-
ment firmly in your right hand.
Tap your right hand on the palm of your left
hand until the battery falls out.
NOTE
When a battery is installed, the pump per-
forms a self-test during which all symbols
will be displayed and the pump will emit
acoustic signals.
Under average operating conditions, each
battery should last about six months.
The pump retains all programmed values.
The data are not lost when the battery is
replaced or removed or if the pump isn't
used for prolonged periods. Programmed
partial volume is the only value that is not
saved if the battery is removed.
The door is fitted with a silicone seal that
must remain in its lodging as shown in
figure.
Silicon seal