Active limitation of  bolus
doses number (not possible
to deliver bolus doses)
Refer to user guide
at page  51
Refer to user guide
at page  50
Refer to user guide
at page  51
Bolus dose not available 
(value set for bolus dose is 0) 
INFUSIONS
Active limitation of interval bet-
ween bolus doses (not possible
to deliver bolus doses)
Keypad  locked
Unlock the keypad
CODE
DESCRIPTION
ACTION
10
20
10
20
10
20
10
20
Alarm messages are accompanied by audible signals.
Re-start
To re-start the device following an alarm message, remove the battery and 
wait at least 10 seconds before re-inserting it. 
*Warning
After Er,8 signalling, and the subsequent initialization, the standard para -
meters would be the factory ones (see page 16).  Therefore, should this
condition take place, you need to re-program the parameters assigned by
the physician.
The parameters programmed by the physician have to be recorded on the
note-page (see page 82).