User guide

SECTION
14
74
MAN 01/06/USA/p CRONO FIVE 11.10
Active limitation of bolus
doses number (not possible
to deliver bolus doses)
Refer to user guide
at page 51
Refer to user guide
at page 50
Refer to user guide
at page 51
Bolus dose not available
(value set for bolus dose is 0)
INFUSIONS
Active limitation of interval bet-
ween bolus doses (not possible
to deliver bolus doses)
Keypad locked
Unlock the keypad
CODE
DESCRIPTION
ACTION
10
20
10
20
10
20
10
20
Alarm messages are accompanied by audible signals.
Re-start
To re-start the device following an alarm message, remove the battery and
wait at least 10 seconds before re-inserting it.
*Warning
After Er,8 signalling, and the subsequent initialization, the standard para -
meters would be the factory ones (see page 16). Therefore, should this
condition take place, you need to re-program the parameters assigned by
the physician.
The parameters programmed by the physician have to be recorded on the
note-page (see page 82).