User guide
A
NNEX 1
85
ACCURACY TEST
MAN 01/06/USA/p CRONO FIVE 11.10
The tests have been carried out in compliance with the standard IEC 60601-
2-24 - Medical electrical equipment – Part 2-24: Particular requirements for
the safety of infusion and controllers. The following graphics show the accu-
racy of the pump during the administration.
1.1 - Start-up flow rate
• Programmed flow rate: 1000 mcl/h