User guide
ANNEX 4
89
ELECTROMAGNETIC COMPATIBILITY
MAN 01/06/USA/p CRONO FIVE 11.10
The tests have been carried out in compliance with the standard IEC 60601-
2-24 - Medical electrical equipment - Part 2-24: Particular requirement for
the safety of infusion pump and controllers.
4.1 – Emission tests
The test has been executed with the antenna in horizontal and vertical pola-
rization.
4.1.1
4.1.2 - Result
No degradation of performance or loss of function is allowed during the
test.
4.2 - Electrostatic Discharge
Discharges in air have been carried out in proximity of the push-buttons, in
proximity of two angles of the display and two angles of the battery com -
partment.
Contact discharges have been carried out on the two eyelets and two points
of the anodized metallic body.
4.2.1
4.2.2 - Result
No degradation of performance or loss of function is allowed during the test.
TEST Coupling port Range frequency
Radiated Emission Enclosure 30 - 1000 MHz
TEST Coupling port Test level
Electrostatic
Discharge
Enclosure
15 kV air discharge, 8
kV contact discharge,
pos. and neg.