User guide

90
MAN 01/06/USA/p CRONO FIVE 11.10
4.3 - Radiated immunity
The A test has been executed with horizontal and vertical field polarization.
The B test has been executed with horizontal and vertical field polarization,
with steps of equal frequency to 1% of the fundamental one.
4.3.1
4.3.2 - Result
No degradation of performance or loss of function is allowed during the test.
TEST Coupling port Frequency range Test level
A - Radiated
immunity
Enclosure 26-80 MHz
10 V/m 80% AM
1kHz
B - Radiated
immunity
Enclosure 80-1000 MHz
10 V/m 80% AM
1kHz