i CONTENTS 1 INTRODUCTION 1 Instrument overview .............................................................................................................. 2 Instrument specifications ....................................................................................................... 3 Safety warnings .................................................................................................................... 3 2 UNPACKING AND ASSEMBLY 5 Unpacking the CMRV-4500 .........................
ii 5 VISCPRO® INSTALLATION (continued) Running the software ........................................................................................................... 20 Loading software .................................................................................................... 20 Displaying the Instrument View window ................................................................... 20 Viewing report data .........................................................................................
iii 9 USING THE CMRV SOFTWARE (continued) CMRV module menu options .............................................................................................. 49 Configure options ................................................................................................... 49 Print Instrument and Tray Settings ........................................................................... 49 Instrument Settings ..........................................................................................
iv 13 ANALYSIS CONFIGURATION OPTIONS 71 Creating an analysis ............................................................................................................ 71 Sorting analysis data ........................................................................................................... 73 Using the date filter ............................................................................................................. 73 Using the sample/error filter ......................................
v 18 USING THE DATABASE MANAGER 101 Archiving old data ................................................................................................. 102 Changing the database directory ............................................................................ 102 Importing archived data ........................................................................................ 103 Repairing/compacting the database ........................................................................ 103 Exit ......
vi This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
1 CHAPTER 1 INTRODUCTION Purpose of the manual This manual has been written to provide the information necessary for proper installation, operation, and maintenance of the CANNON ® MiniRotary Viscometer (CMRV-4500). Instrument utility The CANNON ® Mini-Rotary Viscometer is used to measure the apparent viscosity and yield stress of engine oils and drive line lubricants within the temperature range of -10°C to -40°C using ASTM test methods D 4684, D 3829, D6821 and D 6896.
2 Networking capability for multiple instruments The VISCPRO® software can control/monitor up to four CMRV instruments with one computer via RS-485 serial connections. See APPENDIX D for more information. Instrument overview The CANNON ® Mini-Rotary Viscometer is designed for precision control of temperature over time, enabling accurate yield stress and viscosity measurement of oil samples in conformance with ASTM D 3829, ASTM D 4684, ASTM D 6821 and ASTM D 6896 test methods.
3 Instrument specifications C MR V 4500 SER IES SPEC IFIC ATION S Instrument Model Model 4500 D imensions 317 mm wi de × 260 mm deep × 489 mm hi gh (12.5 × 10.25 × 19.25") Weight 18.
4 • General Caution Hot Surface Caution Refer all service and repairs to qualified personnel. In addition to the warnings previously listed, additional cautions are posted throughout the manual. These warnings may be designated by an appropriate symbol inside an equilateral triangle. General cautions are indicated with an exclamation point (see diagram, left). Read and follow these important instructions.
5 CHAPTER 2 UNPACKING AND ASSEMBLY Unpacking the CMRV-4500 CAUTION Some CMRV components, including the Air/Water Heat Exchanger, are quite heavy. To avoid injury, obtain necessary assistance when lifting and moving shipping cartons and heavier unpacked components. Remove all components from the shipping container(s). Remove any and all packing materials (styrofoam, etc.) from the components. Verify reception of shipped materials by comparing equipment items with packing/parts list(s).
6 Set of weights: One 150-gram weight • One hook-cage • Nine additional weights 1 Bottle of N105B with test sample and data sheet Tubing and tube clamps Plexiglas® Top Cover Instruction & Operation Manual 9 Rotor bearing pins 10 Rotor locking pins User-Supplied Equipment The user must supply an electrical power source matching the electrical requirements indicated on the rear panel of the CMRV-4500 model.
7 CHAPTER 3 CMRV-4500 APPARATUS DESCRIPTION CMRV-4500 unit/accessories The CMRV-4500 unit contains an aluminum block with a heater for warming the block. Nine viscometric cells are closely fitted into nine holes in the block (see photo). There are also two thermom- CMRV-4 unit with rotors inserted eter wells in the block.
8 Rotor crossbar The rotor crossbar is used to hold the loop at the end of the thread. It also serves as an indicator for (optional) manual timing of rotor rotation. Rotor locking pins The rotor locking pins String on rotor and pulley-wheel assembly are used to prevent unwanted rotor rotation. When the locking pin is lowered over the rotor crossbar (see photo), rotation is prevented. When the pins are in the raised (detent) position, the rotors are free to rotate.
9 Hook-weights Hook-weights are hung on the plastic ring at the free end of the thread to apply the required force to the rotor during yield stress and viscosity measurements. Nine disk weights, one hook-cage, and one larger hook-weight are provided. Weights vary depending on the ASTM Method used for testing.
10 Front panel operations CMRV front panel Control features CMRV front panel control features are simple and functional. The central switch on the front panel is the power switch for the unit. The lightemitting diodes (LEDs) above the switch indicate function of the heating and cooling systems. The Sensor connection on the right side of the panel mates with the cable from the pulleywheel optical sensor.
11 CHAPTER 4 SETUP The instructions in this chapter are for setting up a single CMRV-4500 with the Air/Water Heat Exchanger. For additional information on multiunit configurations, see APPENDIX D. CAUTION Do not apply power to the CMRV-4500 or Heat Exchanger until all connections are secure. Physical placement CAUTION Some CMRV components, including the Air/Water Heat Exchanger, are quite heavy. To avoid injury, obtain necessary assistance when lifting and moving these components.
12 Tubing connections There are two tubing connections on the rear of the CMRV-4500 housing (see diagram, next page). The fitting on the left is the coolant INLET (for introducing coolant to the CMRV-4500) and the fitting on the right of the rear panel is the coolant OUTLET (for returning coolant to the Ex changer).
13 Exchanger when the unit commences operation. You will be able to observe the liquid circulating inside the reservoir from the opening at the top of the Exchanger during normal operation. If air bubbles are consistently visible in the coolant lines, add additional mixture until the Exchanger is full. CAUTION The water-antifreeze mix should be replaced annually for reliable performance and to prevent corrosion of internal components (see Flushing and draining the Air/Water Heat Exchanger, this chapter).
14 Heater Insert the heater plug from the upper section of the CMRV-4500 housing into the HEATER receptacle on the rear panel of the CMRV controller. CMRV-4500 power cord Make sure the CMRV-4500 power switch is in the OFF position.
15 CAUTION Make sure that the assembly is seated securely on the track. 3. Replace the left front screw. This Pulley-wheel assembly mounted on slide track will prevent the assembly from sliding off the track. 4. Plug the free end of the pulley-wheel sensor wire into the jack labelled WHEEL on the front of the CMRV-4500 controller. NOTE To disconnect the sensor, pull it out by the knurled portion of the plug.
16 Cleaning the instrument housing CAUTION Before cleaning the CMRV housing, turn off the instrument and unplug the power cord. Do not clean the instrument unless the cell temperature is within 10°C of ambient. Periodically clean the outside of the unit with a damp cloth moistened with water and/or a mild detergent solution. Flushing and draining the Air/Water Heat Exchanger The water-antifreeze mixture should be replaced annually for reliable performance and to prevent corrosion of internal components.
17 7. Replace the connector on the Exchanger hose, and reattach the connector to the fitting on the Exchanger (see photo). Draining fluid from the Exchanger After the antifreeze mixture has been flushed from the Air/Water Heat Exchanger per the preceding procedure, drain the system as follows: 1.
18 4. Replace the external housing on the Air/Water Heat Exchanger and secure it with the six screws previously removed. If necessary, reseat the gasket around the reservoir opening with a small screwdriver. 5. Add water-antifreeze mixture to the Exchanger per the earlier instructions in this chapter to complete the water-antifreeze flush and drain procedure.
19 CHAPTER 5 VISCPRO® INSTALLATION VISCPRO ® for Windows ® XP® VISCPRO® is a powerful new software product providing a generic instrument interface for controlling and operating your CANNON® instrument(s) via computer. VISCPRO® also includes reporting/analysis modules for processing and displaying sample data. Installing VISCPRO ® software To install the VISCPRO® software, follow the instructions below in the sequence presented.
20 Running the software Loading software Make certain that your CMRV instrument is properly connected to your computer and the CMRV power switch is ON. Then start the VISCPRO® software by double-clicking the VISCPRO® icon on your Windows® desktop. Or click Start/Programs/VISCPRO/VISCPRO.EXE).
21 The Instrument View window NOTE To display the Instrument View window, click View Instrument from the Main menu. The View Instrument window will appear. Then click the MRV instrument ID to display the list of available configurations. For now, select the default configuration and click OK. Instrument ID/Type Instrument Configuration If the Available Instruments list box is blank, your instrument(s) may not be on-line. Check cable connections and make certain the instrument power switch is ON.
22 Viewing report data After you have completed CMRV calibration and testing (see next chapters), you will be able to display a report window with CMRV-4500 test results. To access the database and display test data, click Analyses/ View Analysis.
23 Checking Configuration data Follow the procedures in the next several sections of this chapter to select and verify the instrument and calibration settings to ensure that they conform to the actual characteristics of your CANNON® instrument. Configuration protection To check the configuration settings for your instrument(s), you must log in to the security system as a manager. The software is installed with a default Manager account.
24 The Instrument Settings window 4. Use the ID field to input instrument identification information using up to 30 alphanumeric characters. The remaining fields in the Instrument Settings window are noneditable (information is obtained via serial communication with the CMRV instrument). The Model: field will indicate the model of your instrument. The S/N: field indicates the serial number from the label on the rear service panel.
25 NOTE Some fields/options are security-protected, and may not be viewable. The View Cell Constants window 3. Click Configuration/Print Instrument Settings to open the Windows® Print window. Select the desired printer and click OK to print current calibration settings. NOTE CANNON recommends printing calibration settings each time the calibration values change.
26 features required for testing samples and performing other routine operations. Managers may access advanced configuration and reporting options. The Service level permits access to all instrument and security parameters. Initial security setup The VISCPRO® software installation creates a security account for one manager with a blank password. After VISCPRO® installation, a password should be assigned for the manager using the Change Password menu option from Main.
27 Procedure 1. Click Log In from Main. 2. Select your Manager or Service level user name from the User Name list box. Input the correct password in the appropriate field. 3. Click OK. 4. Click Update User Information from Main. 5. Select the desired user from the User Name list box. NOTE To delete a user, just click Remove User after selecting the user name. The account will be immediately and permanently removed. 6.
28 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
29 CHAPTER 6 PREPARING FOR CMRV TESTING/CALIBRATION The procedures in this chapter should be followed when preparing for CMRV-4500 testing or calibration. NOTE In the event that the protocols of your ASTM test method differ from any of the procedures outlined in this manual, the ASTM method should take precedence. A/W Heat Exchanger To prepare for operation of the CMRV-4500 unit, turn on the POWER switch on the upper rear panel of the Air/Water Heat Exchanger.
30 CAUTION! Exposure to acetone liquid or acetone vapor may damage the cover. 2. Turn on the CMRV-4500 (the green power indicator on the switch will light and the left panel lights will flash once). NOTE The CMRV cleaning procedure may be initiated from a "cold" CMRV; however, a calibration or profile should not be initiated until the instrument has completed a warmup period of approximately 45 minutes.
31 8. Wait for the CMRV-4500 to reach the cleaning temperature. NOTE When the CMRV-4500 cleaning cycle is activated, the internal heater raises the viscometer block temperature 2-3°C per minute to approximately 50°C. Precise temperature control is not necessary for cleaning. 9. When the CMRV-4500 has reached cleaning temperature remove the rotors and wipe excess oil from them. Gently place the rotors into a solvent-resistant container, such as a 500-ml glass beaker.
32 17. When you have finished cleaning the viscometer cells and rotors, click Stop Cleaning Cycle. Then click to exit the Cleaning ... window. NOTE After the cleaning cycle, you should permit the cells and rotors to dry for 15 minutes before introducing oil sample into the viscometric cells. Inserting rotors The viscometric cells are numbered from left to right when looking at the viscometer from the pulley-wheel side, with cells 1, 3, 5, 7, and 9 in the front row and cells 2, 4, 6, and 8 in the back row.
33 CHAPTER 7 CALIBRATING THE CMRV TEMPERATURE PROBE The CMRV temperature probe must be calibrated before performing the initial cell calibration or running temperature profiles. A complete calibration session involves setting temperature offsets in the VISCPRO software for key temperature checkpoints beginning at 80°C, then 50°C, then other temperatures at 5-degree increments from 0 to -40°C.
34 NOTE In each calibration session, the initial probe calibration must be performed at 80°C (see procedure below) before calibrating at other temperatures. When prompted by the VISCPRO software: "Do you want to zero the offset at 80°C before calibrating?", click YES the first time in the session that you attempt to calibrate at 80°C. Thereafter, click NO unless you would like to start the calibration over from the default offset value.
35 CHAPTER 8 CALIBRATING THE CMRV CELLS Initial calibration To ensure the accuracy of sample data, the CMRV-4500 cell calibration procedure must be completed after initial installation of the software. First complete the temperature probe calibration (see previous chapter) and instrument preparation (Chapter 6). Then follow the procedure in this chapter to calibrate the CMRV cells.
36 η = t3r = viscosity of the calibration standard (cP) time for 3 revolutions of the rotor (sec) Cell calibration procedure Calibration standards CANNON ® viscosity standard N105B is recommended for calibration at -20°C and -25°C. This calibration is required for running ASTM profiles included with the software. At -20°C, N105B has a nominal viscosity of 30,000 cP (mPa·s). At -25°C, N105B has a nominal viscosity of 56,600 cP (mPa·s).
37 wound around the rotor shaft. Then pass the thread below the crossbar and continue turning the rotor clockwise (as viewed from above) to wind 20 closely spaced turns of thread around the shaft below the crossbar without overlapping the turns. About 200 mm (8 inches) of thread will remain.
38 NOTES To reduce the chance of separating the mercury column during the heating phase of the temperature profile, you may wish to initially remove the calibrated thermometer from the CMRV unit. Store the thermometer in a vertical position. You may replace the calibrated thermometer when the CMRV has cooled to 30°C or lower.
39 7. Remove the Plexiglas® cover and align the pulley-wheel with the cell to be calibrated. 8. Place the thread from this cell over the pulley-wheel. Hook the calibration weight specified by the Method to the end of the thread (see photo). 9. From the Calibrate window, click the desired cell number for calibration. 10. Press the blinking Start Test button on the CMRV controller. 11. Raise the locking pin quickly and smoothly. The weight will apply torque to the rotor.
40 16. When acceptable values have been obtained, click Abort Temperature Control. Then click Save Data to save the new calibration constants. Do not omit this step or new calibration data will be lost. 17. Close the Calibrate Cell Constants window and then click Main/ Save Instrument. The Save Instrument window will appear. 18. In the Save As box, input a name for the new instrument configuration (which includes the newly-calculated calibration constants). Then click OK to save the configuration.
41 CHAPTER 9 USING THE CMRV SOFTWARE VISCPRO ® generic instrument interface Your software for Windows® 95/NT® is comprised of a generic instrument interface (VISCPRO®) and a collection of instrument/analysisspecific modules. This chapter of the manual will explain the software options for the VISCPRO® program and other modules commonly bundled with the instrument software. Additional software modules, customized software, and any documentation for add-on software options, are provided separately.
42 Main options The View Instrument option opens the View Instrument window which permits you to view existing instruments and configurations. You may also select and restore elements from saved configurations for on-line instruments. The View Instrument window The View Instrument window is comprised of: Two list boxes (Available Instruments and Saved Configurations) Restore ...
43 Saved Configurations: The Saved Configurations list box (located directly under the Restore . . . check boxes) allows you to make changes to the current configuration for any networked CANNON® instrument by restoring any or all elements (Instrument, Tray or Sample) of a saved configuration. To restore saved configuration settings, select the instrument from the Available Instruments window and then click on the desired configuration from the list of saved configurations.
44 The Save Instrument option opens the Save Instrument window. The Save Instrument window permits you to save all current instrument information for any selected instrument, including Instrument Settings, Tray Settings and Sample settings, to the database. Once saved, the instrument information may be restored at any time using the Restore options from the View Instrument window.
45 operations. Managers may access advanced configuration and reporting options. The Service level permits access to all instrument and security parameters. Initial security setup The VISCPRO® software installation creates a security account for one manager with a blank password. After VISCPRO® installation, a password should be assigned for the manager using the Change Password menu option from Main. Managers may add or change accounts for technician status personnel.
46 2. Select your Manager or Service level user name from the User Name list box. Input the correct password in the appropriate field. 3. Click OK. 4. Click Update User Information from Main. 5. Select the desired user from the User Name list box. NOTE To delete a user, just click Remove User after selecting the user name. The account will be immediately and permanently removed. 6. Enter and verify the desired password, and select user security status using the appropriate list boxes. 7.
47 Each analysis generates a report based on instrument data in the VISCPRO® database. The following analyses are shipped with the software: Analysis types Error Data Export, MRV Data Export analyses—configures data for output to a file, parallel port, or serial port. Error Log Table—lists error messages and related data. MRV Data Table—displays test information.
48 Select the desired analysis from the directory tree. After an analysis has been selected, you can configure the analysis to display the information you want from the central VISCPRO® database. The Choose Analysis window provides three options for doing this: Click the appropriate radio button .
49 windows, click Window from the VISCPRO® primary display and select the desired instrument view. Restoring a saved window configuration does not affect current CMRV sample testing operations. CMRV module menu options In addition to the VISCPRO® menu functions, unique software application modules for each type of CANNON® instrument generate additional interface options. The modules determine the characteristics, function and appearance of VISCPRO® software menus and windows.
50 Instrument Settings window button options: saves the current instrument settings and exits the Instrument Settings window. closes the Instrument Settings window without saving any configuration changes. Profile Designer The Profile Designer option opens the Profile Designer window, enabling selection and editing of temperature profiles. For information using the Profile Designer and Profile Editor, see Chapter 14, Designing Customized Profiles.
51 4. Click on the desired cell number (or select it from the drop-down box). 5. Input the desired value(s) for the calibration(s) and the viscosity of the calibration standard. Constant 1 represents the most current calibrated value, and Constant 2 represents the value from the previous cell calibration). The viscosity of the standard is indicated on the bottle label. NOTE If constants are edited with invalid information, data from samples tested using the constant information will be compromised. 6.
52 Restore procedure To restore instrument settings from a saved configuration: 1. Click on View Instrument from the Main menu options. 2. Select the desired instrument from the left list box. 3. Select the desired saved configuration from the right list box. 4. Use the check boxes to select the desired configuration elements to be restored. Restore Instrument Settings restores instrument properties from the saved configuration. Restore Tray Settings is not applicable to the CMRV instrument.
53 Temperature: Click Mass: Click to select the desired temperature for calibration. to select the desired test mass for calibration. Known Viscosity: Enter the known viscosity of the standard, as printed on the standard bottle label. Calibrate ... : After you have entered the calibration data described above, click the Calibrate ... button to access the CMRV Cell Calibration window.
54 Service menu options To access the service options, click Service from the VISCPRO® primary display and select the desired instrument. MRV Monitor The MRV Monitor window permits the user to view current operational values for the CMRV instrument. These settings may be helpful to CANNON ® technical personnel when troubleshooting difficulties with the CMRV. Cleaning Service Select Cleaning Service to open the Cleaning Service window.
55 CHAPTER 10 SUMMARY OF TEST PROCEDURE In the ASTM D 3829, ASTM D 4684 and ASTM D 6821 test methods, a 10-ml sample is placed in a test cell, pre-heated to a specified temperature, and held at that temperature for a specified time to ensure complete solution of all components of the oil. The temperature of the cell is then lowered to the test temperature at a programmed cooling rate.
56 NOTE Most commercial oils do not have a yield stress when tested as specified by the SAE J300 Viscosity Classification. For the complete, definitive description of the ASTM D 4684 method, see the provided Standard Test Method for Determination of Yield Stress and Apparent Viscosity of Engine Oils at Low Temperature, Designation: D 4684, also available in the Annual Book of ASTM Standards, Volume 05.03. NOTE Nothing in this manual is intended to supersede the provisions of the ASTM test method.
57 The temperature profile for this method consists of a 1.5-hour soak at 50°C ± 1°C, a 2-hour nonlinear cool-down to the test temperature, and a 14-hour soak ± 0.02°C at the test temperature. The entire temperature profile, up to the point of measuring yield stress and viscosity, requires 17.5 hours.
58 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
59 CHAPTER 11 RUNNING PROFILES NOTE The guidelines in this section are not intended to supercede any test method. Always refer to the method for validation of any test procedure. Test preparation To run a temperature profile, first prepare the CMRV-4500 for testing (see Chapter 6). Make sure that the Air/Water Heat Exchanger POWER switch is on and that the unit is operating normally. Procedure 1. Place 10 ml of test oil in each of the viscometric cells to be used.
60 Winding rotors 7. To wind the rotors in preparation for the profile, slide the pulleywheel assembly along the track until the pulley is aligned with the first viscometric cell. Use the engraved lines on the slide bar as a guide to determine the proper position of the left side of the pulleywheel block. 8. Place the loop of a thread over the left end of the crossbar at the top of the rotor shaft. 9.
61 cover (Step 9.2.6 in the ASTM D 4684-98 method and step 9.1.6 in the ASTM D 3829 method). The tubing may be removed, along with the cover, prior to viscosity/yield stress testing. 14. Whatever the methodology, the purge should continue throughout the temperature profile using dry gas at a flow rate between 20 and 30 milliliters per minute. Starting a profile 1.
62 5. At the conclusion of the temperature profile you are ready to test samples. See Chapter 11 for the test procedure. Profile operation notes Instrument Status window During the profile, the Instrument View window will display the current status of the profile, the current test temperature, the desired temperature, time elapsed, and the time remaining until the yield stress and viscosity tests should be run. Display data is normally updated every 0.5 seconds.
63 Adjusting temperature at the end of a profile In the event that the CMRV instrument fails to attain the desired temperature at the conclusion of a profile, it is possible to adjust the instrument temperature using the following procedure: 1. Wait until the profie is completed and the Temperature Correction button is active. 2. Click in the Thermometer Temperature field and type the current temperature (as indicated by your calibrated thermometer) to the nearest 0.
64 Zooming in To view a section of the profile graph in greater detail, click and drag (press and hold down the left mouse button) the cursor to "draw" a box around the desired graph area. Then release the mouse button. The graph will be redrawn to include only the selected area. Zooming out To reduce the graph magnification, click the Refreshing the graph If you are displaying the profile graph while a temperature profile is running, you may not be viewing the most current temperature data.
65 CHAPTER 12 MEASURING YIELD STRESS AND VISCOSITY NOTE The guidelines in this section are not intended to supercede any test method. Always refer to the method for validation of any test procedure. Yield Stress and Viscosity testing should take place shortly after the profile is completed. NOTE ASTM D 4684 requires completion of yield stress and viscosity tests within 30 minutes of the conclusion of the profile.
66 ASTM D 3829 and 4684 specify an initial weight of 10-grams, and increase of weight in 10-gram increments. ASTM D 6821 specifies an initial weight of 2.5 grams and increase of weight in 2.5-gram increments. 6. Ensure that a Sample ID has been entered for each sample. Then click for the desired sample. OR use the Tab key to highlight the yield stress icon and then press R. The Yield LED (lightemitting diode) on the CMRV-4500 will glow and the Start Test button will begin flashing rapidly. 7.
67 10. If testing viscosity for the sample, follow the directions for viscosity testing (see Measuring Apparent Viscosity, next page). 11. Repeat steps 1-10 to perform yield stress/viscosity tests for all remaining cells to be tested in order from left to right. Notes on yield stress testing Step 9 (see previous section) is required for ASTM D 3829. The ASTM D 4684 method is a pass-fail method for fresh oils.
68 5. Press the flashing Start Test button and proceed immediately to step 6. 6. Lift the locking pin and release the weight by lowering your hand quickly and smoothly. The CMRV-4500 will measure the speed of rotor rotation by timing the revolution(s) of the pulleywheel as the weight falls. NOTE The total number of rotations measured by the CMRV-4500 during the viscosity test depends on the viscosity of the sample and the values chosen in the Advanced Settings box.
69 Notes on ASTM methods The protocols for measuring viscosity differ considerably between the ASTM methods. For further information on the methodology used to measure viscosity, consult the applicable ASTM method. Printing yield stress/viscosity test results To print the CMRV data form, including viscosity and yield stress test results: 1. Open and/or configure the appropriate analysis. 2. Use the mouse to highlight the desired data. 3. Click Print from the Main menu options. 4.
70 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
71 CHAPTER 13 ANALYSIS CONFIGURATION OPTIONS Data obtained from all instruments during sample testing is stored in the central VISCPRO® database. To view data, you must create an analysis configuration requesting the desired sample information in the desired format. Analysis configurations can be saved and later restored. The analysis configuration options provide powerful tools for reporting sample information.
72 4. Click on the radio button action: corresponding to the desired analysis The Open selected Analysis option will open the selected analysis without providing an opportunity for modification of the analysis. The View and Open Existing Configuration option will permit the user to view/modify an existing analysis (see notes and procedure following). The Define New Analysis option will create a new analysis (see notes and procedure following).
73 Sorting analysis data After you have displayed an analysis, you may sort alphanumerically by any of the table headings (the default sort for most reports is Time Stamp). To sort, simply click on the desired column heading. The table will be displayed with the new sort order. NOTE The Sort function is only available for an analysis when the Dynamic Update mode option is NOT selected.
74 To automatically update the report with sample data as tests are completed, click the Use Dynamic Update radio button. Click the Do not show previous samples box if you do not wish to display samples tested prior to the time the analysis is opened. Use the spin controls to select the Sample Time Window (the desired range of recent samples to be included in the analysis). Data from samples tested within [x] hours will be displayed in the analysis.
75 Then select the desired instruments and profile names from their respective list boxes. To select a range of profile names/instruments, click on the first desired item and then hold down the S key while clicking on the last desired item in the sequence. To individually select profile names/instruments from the list boxes and add them to the analysis, hold down the C key while clicking on each desired profile/instrument.
76 5. You will be prompted to save the new analysis configuration. After you make your selection(s), the analysis will be performed and displayed using the new configuration settings. Resizing table columns To resize columns from a displayed analysis for easier viewing/printing, move your mouse pointer/cursor to the edge of the desired column heading. The mouse pointer will change to a bi-dimensional arrow. Click and drag the edge of the heading to the new location and then release the mouse button.
77 print options from the Windows® Print window and click OK to print the analysis. (Consult your Windows® manual and printer/print driver documentation for more information on print settings). It is also possible to select individual or multiple samples from an analysis for printing (see next section). NOTES The font size for printed analyses will be automatically adjusted to fit data columns to the selected printer paper size.
78 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
79 CHAPTER 14 DESIGNING CUSTOMIZED PROFILES Cooling Profiles The cooling profile may be pictured graphically as a curve (temperature over time) calculated from up to 25 individual user-defined data points. To design a profile, you will use the Profile Editor from the VISCPRO® Profile Designer window to input time, temperature and (if desired) tolerance values for each point: Time: The time during the sample- run at which a desired temperature is to be achieved. Acceptable values for hours are 0- 99.
80 Opening the Profile Designer To create a new profile or modify an existing profile, Click Configure/ MRV/Profile Designer. The Profile Designer window will open. The Profile Designer window The Profile Designer window provides button options for copying and modifying existing profiles or creating entirely new profiles.
81 the profile library. The Profile Designer window also displays Yield Stress and Viscosity test parameters for the profile. The Profile Designer interface options are explained briefly in the following section: Interface options The profile library The profile library consists of two types of profiles—ASTM-defined profiles and User-defined profiles. Click the "down" arrow ( ) to display the list of available profiles.
82 Click the Delete Selected Profile button to delete the currently-displayed User Defined Profile (ASTM Standard Profiles cannot be deleted from the Library).
83 NOTE ASTM-defined profiles may be selected and viewed but may not be altered. To make revisions to an ASTM-defined profile, you must first Copy Selected Profile ... and then Save it as a User-defined profile. 1. To define the cooling profile, input desired information (Hours, Minutes, Temperature and Tolerance) in the appropriate fields for a single point in the temperature profile. Adding a data point to the profile 2. Click the Add Point button to add the data to the profile.
84 TO delete the point click Yes. To keep the selected point(s) click No. Cooling profile limitations CMRV cooling capacity The CMRV cooling capacity changes proportionally with the temperature at which the instrument is controlling. Greater cooling capacity is achieved at warmer temperatures, and capacity decreases as temperature decreases. This makes linear rates-of-cooling impossible in relatively short time frames.
85 NOTE When the user is designing an entirely new profile “from scratch”, the test parameters will default to ASTM D 4684 specifications. Copies or modifications of standard profiles will default to the parameters associated with the ASTM method (D 3829 or D 4684). Changing test parameters To change test parameters: 1. Open the Profile Designer. 2. Select the desired profile and click the Create ..., Copy ... or Edit ... button as desired. The Profile Editor window will open. 3.
86 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
87 CHAPTER 15 MRV DATA TABLE ANALYSIS The MRV Data Table analysis is designed to permit convenient viewing of data collected from samples which have been tested with the CMRV. The sample analysis displays sample data in a tabular format.
88 Configuring the MRV Data Table For general information on using analyses, see Chapter 13. To access and configure the MRV Data Table, follow the procedure below: 1. Select Analyses from the VISCPRO® primary menu options. 2. Select View Analysis ... from the Analyses menu. The Choose Analysis window will appear. 3. Double-click Basic Package (or click the adjacent “ ”sign) from the list of Available Analysis Packages. 4. Select MRV Data Table from the Basic Package report options.
89 Date Filter—allows you to select date/time parameters for the analysis (see Chapter 13 for additional details). Sample Filter—allows you to select which tests will be included in the analysis (see Chapter 13 for additional details). MRV Report Filter—allows you to select what sample data will appear in the Sample Data Table and how the data will be displayed 7.
90 8. When you have completed the configuration, click OK. The program will prompt you to save the configuration. 9. Click Yes to save the configuration. The Save Configuration window will appear. The Save Configuration window 10. Type the name of the new configuration in the Save As: field. Or click the name of a preexisting configuration in the Existing Configurations list box to replace the existing configuration with the new configuration settings. 11. Click OK.
91 CHAPTER 16 ERROR LOG TABLE ANALYSIS The Error Log Table is designed as a troubleshooting tool to display error messages generated by the software during automatic processing of sample data. The Error Log Table displays data in a tabular format.
92 Choosing the Error Log Table analysis 4. Click on the Define and Open New Configuration radio button (or verify that the option is selected). NOTE If you have already configured and saved an analysis, its name will appear in the list box on the right side of the window. If you click on an existing configuration and click OK, the analysis will be performed using the selected configuration settings. It will not be necessary to complete the remaining steps in this procedure. 5. Click OK.
93 8. Click Yes to save the configuration. The Save Configuration window will appear. The Save Configuration window 9. Type the name of the new configuration in the Save As: field. Or double-click the name of a preexisting configuration in the Existing Configurations list box to replace the existing configuration with the new configuration settings. 10. Click OK. The analysis will be performed and displayed using the selected configuration settings.
94 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
95 CHAPTER 17 EXPORT ANALYSES The VISCPRO® Export analyses (MRV Data Export and Error Data Export) provide a convenient operator interface for configuring sample or error information from the sample database for serial output and exporting it in ASCII text format. The three port output filters (Date Filter, Sample Filter/Error Page and Port Output Format) permit the user to select and output desired data to a file, LPT port or serial port.
96 ASCII Code—An ASCII character code input by the user, as defined by its byte code Smpl Type—the type of sample Visc Time—Time required for viscosity test completion Visc Rev.—Number of revolutions during viscosity test Cell Const—Cell constants associated with the test Mass—Test weight (in grams) Cal Temp—Cell calibration temperature Cal Visc—Viscosity of calibration standard Mass Inc.
97 Choosing an analysis 4. Select the desired Data Export analysis option. The CMRV Ports Output Configuration window will open. 5. Click the Define and Open New Configuration radio button define a new analysis configuration. NOTE to If you have already configured and saved an analysis, its name will appear in the list box on the right side of the window. If you click on an existing configuration and click OK, the analysis will be performed using the selected configuration settings.
98 Filters a. Complete selection of Date and Sample/Error filter options per the instructions in Chapter 13. Ports b. Click the Port Output Format tab and click Add Port from the button options to open the Select Port window. Select the desired serial port(s) and/or files for output and verify the configuration settings for each. Then Click OK. Added ports will be displayed in the port list box. NOTE If you select NEW FILE for output, click the button to open the Windows Save As: box.
99 The Port Output tab Adding a header g. If you would like to include a header at the beginning of the analysis, click the Header button and add the desired text string via the keyboard. Format the entry as desired using the Carriage Return (CR) and Line Feed (LF) options as necessary to indicate line breaks. Then click OK. Adding a footer h. If you would like to include a footer at the end of the analysis, click the Footer button and add the desired text string via the keyboard.
100 9. If you do not wish to save the configuration, click No. The analysis will be displayed and the data will be sent to the selected ports. If you wish to save the configuration, click Yes. The Save Configuration window will appear. Type the name of the new configuration in the Save As: field. Or double-click the name of a preexisting configuration in the Existing Configurations list box to replace the existing configuration with the new configuration settings. Then click OK.
101 CHAPTER 18 USING THE DATABASE MANAGER The Database Manager is a separate software program which is automatically installed in the same directory as your VISCPRO® software.
102 Periodic archiving CANNON® Instrument Company recommends periodic archiving of sample data to limit the size of the operating database and enhance the software response speed when generating analyses and accessing the database. Starting the program To start the Database Manager software: 1. Exit the VISCPRO® application. Then click on the Windows® Start bar. 2. Select Programs/VISCPRO/VISCPRO Database Manager from the list of options.
103 location to another. Rather, it allows the user to select the location for the working database. The database must first be archived (see previous page) or copied to the desired location BEFORE using Change Database Directory. Change . . . procedure NOTE 1. Select Change Database Directory from the button options. A Browse … window will permit you to select the correct location for the database (SAMPLES.mdb) file. The current database directory is indicated at the top of the window. 2.
104 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
105 CHAPTER 19 WARRANTY/RETURN INFORMATION Products limited warranty In addition to other manufacturers’ warrantees, CANNON® Instrument Company (“the Company”) warrants all products (other than reagents and chemicals) delivered to and retained by their original purchasers to be free from defect in material and workmanship for one year from the date of the Company’s invoice to the purchaser.
106 Before returning a CANNON® product for repair or service, make every attempt to identify the problem. If, after careful checking, the problem remains unidentified or unsolved, telephone CANNON® Instrument Company (or the local service agent) to consult with a product specialist. If the specialist cannot recommend a simple solution or repair, CANNON® will authorize the return of the product through the issuance of a Return Authorization number (RA).
107 CHAPTER A APPENDIX A — TROUBLESHOOTING Instrument status window not updating Check serial cable connections. If the status field in the Instrument View window indicates the CMRV is idle, then give the command to initiate the desired operation. If the temperature control lights on the controller are lit or blinking, close the Instrument Status window and reopen it. If temperature control lights on the controller are not lit or blinking, you may have a controller problem.
108 Check the level of the water/antifreeze mixture in the exchanger reservoir and add more mixture if necessary. Also check for pinched hoses or obstructions in the hoses/connectors. If necessary, detach the hoses per manual instructions, remove any obstructions, and reconnect the hoses. If these steps do not solve the problem, the relay in the controller or the Air/Water Heat Exchanger may be defective. Yield stress or viscosity test results inconsistent.
109 CHAPTER B APPENDIX B — REPLACEMENT PARTS LIST Part/Catalog No.
110 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
111 CHAPTER C APPENDIX C — THERMOMETRY Kinematic viscosity and temperature Kinematic viscosity is an extremely temperature-sensitive measurement a change of 1°C can sometimes lead to a viscosity change of 10 percent or more. Therefore, it is not surprising that temperature measurement and control are the most common problems encountered by laboratories performing accurate kinematic viscosity measurements.
112 NOTE Different thermometers have different installation requirements. Refer to the information included with the thermometer in use for specific installation instructions. ASTM thermometer tables The following tables show the ASTM thermometers available from CANNON® Instrument Company: CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
113 NOTE International shipments may be subject to special shipping regulations. ASTM D 445 — Checking the ice point Frequency To achieve an accuracy of ± 0.02°C for calibrated kinematic viscosity thermometers, a check at the ice point must be made. New thermometers should be checked monthly for the first six months, then once every six months. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
114 Method The following text outlines procedures for checking the ice point of a thermometer. The text is adapted from: 1994 Annual Book of ASTM Standards, Volume 05.01, Method E77 ASTM Method E77 contains a detailed procedure for the measurement of ice points. The instructions listed here are specifically designed for the mercury-in-glass “kinematic viscosity” thermometers described in Table 2, and may not apply to other thermometers.
115 NBS Monograph150: Joining separated mercury columns The following text outlines procedures for joining separated mercury columns in thermometers. The text is adapted from: NBS MONOGRAPH 150 Liquid-In-Glass Thermometry Wise, Jacquelyn A. NOTE Many inquiries are received concerning separated mercury column which occur especially during shipment. Since no means of avoiding such occurrences has yet been found, some directions for joining mercury may be helpful and are described below.
116 Organic liquid procedures The procedure for thermometers containing organic liquids is similar. Separated liquid in the stem can be vaporized and permitted to drain down the capillary. Another method consists of gently tapping the stem above the separation against the palm of the hand, forcing the organic fluid to break away from the wall of the capillary and flow down the bore to join the main column.
117 CHAPTER D APPENDIX D — MULTI-UNIT CONFIGURATION Introduction The CMRV 4000 Series instrument have been designed to permit multiunit configurations of up to four CMRV units. Configuration is accomplished via a “daisy-chain” technique using RS-485 connections. A multi-unit interface kit containing all required items for multi-unit configuration is available from CANNON® (Catalogue # 9728-R40 for 120V units; #9728-R45 for 240V units).
118 3. Plug the converter’s RS-232 25-pin connector into the COM port you selected for multi-unit configuration during installation of the CMRV software. If you are using COM 1, a DB-25 to 9-pin adaptor may be required. 4. Select a CMRV unit as the first in the “daisy chain” and attach the three wires from the converter to the ABG connection marked IN on the RS-485 port on the back of the first CMRV chassis as follows (see diagram): Attach the red wire to IN-A. Attach the white or brown wire to IN-B.
119 CHAPTER I INDEX A C (continued) Advanced settings changing 85 Air/Water Heat Exchanger 11, 29 adding water/antifreeze mixture 12 Analyses 46 Analysis types 47 creating 71 deleting 76 menu options for 47 printing 76 printing selected data from 77 reconfiguring 75 resizing table columns for 76 saving 76 sorting data in 73 using the date filter for 73 using the report filter for 75 using the sample filter for 74 viewing 47 wildcard filter options for 75 antifreeze mixture 16 Apparent viscosity 1, 55 A
120 H R (continued) Header adding to analysis 99 Heater 14, 31 HELP software 41 Ice point 29 Instrument settings 49 Instrument Settings window 24 Instrument View window 21 Restore Instrument Configuration 42, 43, 51 Restore Sample Information 43, 52 Restore software options 43 Rotor description 7 inserting 32, 36 locking pins 8 winding 59 Rotor shaft 36, 60 Rotors 29 RS-232 14 L S LF coding in analyses 99 Locking pins 8 S/N: 24 SAE J300 1, 57 Sample Data Table 22 Sample filter 89 sample-run graph 6
121 T (continued) TP-1 2, 55 Tubing connections 12 V View Analysis 47 View and Open Existing Configuration 48 View Instrument 42, 51 window options for 43 Viscometric cells 32 Viscosity 35, 65 viewing data from test 65 Viscosity measurements 67 VISCPRO® software 41 installing 19 W Weights 9 adding 66 Winding rotors 36 Window options 48 Y Yield stress 1, 55, 56, 65 measuring 65 CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.
122 This page intentionally left blank. CANNON® Mini-Rotary Viscometer CMRV-4500 Instruction & Operation Manual Version 1.