Owner's manual

10
2-2-2. Connecting the tube linesD155 with Peristaltic Pump
1) The sampling inlet and outlet ports (1) and (2) are Luer friction-type fittings. See image below for
connecting the tubing.
2) Insert and press the first Connection Tube with Union into the sample inlet Port (1) of the D155 as
shown below. The free end of this tube will later be inserted into the sample container.
3) Connect the second Connection Tube with Union to the sample outlet Port (2). The free end of
this tubing is then connected to the peristaltic pump IN (upper port) fitting as shown below. Connect
the Drain Tube to the peristaltic pump OUT fitting. The free end of the Drain Tube should then be
placed in a suitable waste container such as a beaker or bottle.
2-2-3. Connecting the tube linesD155 without Pump
1) The sampling inlet and outlet ports (1) and (2) are Luer friction-type fittings. See image below for
connecting the tubing.
2) Use a syringe with friction Luer fitting to inject sample into the sample inlet Port (1) noted as “syringe”
of the D155 as shown below.
3) Insert and press the Connection Tube with Union into the sample outlet Port (2) on the D155 and
place the other end of the tube in a suitable waste container such as a beaker or bottle.
OUT
IN
2
1
2
Syringe
CANNON
®
D155 Density Meter Instruction & Operation Manual
Version 1.1b – August 2014; CANNON® Instrument Company
2139 High Tech Road State College, PA 16803 USA