miniAV-X Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.0b— March, 2012; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
9. If you do not wish to save the configuration, click No. The analysis
will be displayed and the data will be sent to the selected ports. If
you wish to save the configuration, click Yes. The Save Configura-
tion window will appear. Type the name of the new configuration in
the Save As: field. Or double-click the name of a preexisting
configuration in the Existing Configurations list box to replace the
existing configuration with the new configuration settings. Then
click OK. The saved analysis will be displayed using the selected
configuration settings, and the data will be sent to the selected ports.
NOTES
You may click Cancel from the Save Configuration window to exit without
saving configuration changes.
The Port Output Analysis cannot be reconfigured. This avoids duplication
of output data for data collection devices your facility may have in place.
You may still view configuration options for a displayed analysis by
selecting Configure Analysis from the Analyses menu and choosing the
correct analysis.
Re-sending export data To re-send Port Output Analysis data for a displayed analysis, first save
the analysis by clicking Analyses/Save Configuration, selecting the
desired analysis, typing the analysis name in the Save As: list box and
clicking OK. Then close the Port Output Analysis window and re-select
Port Output Analysis by clicking Analyses/View Analysis from the primary
menu options. Then click on the desired configuration from the list of
saved configurations and click OK.
Exporting specific sample data To re-send data from specific samples, use Sample Filter options,
including wildcard characters if desired, to reconfigure the analysis to
send only the necessary data.
Port selection Make certain that you have selected the desired port for configuration by
clicking on the port name in the Add Port list box prior to selecting output
data and formatting options for that port. OUTPUT FOR EACH ADDED
PORT MUST BE CONFIGURED SEPARATELY.