Viewing/editing setup information
NOTE
For first-time installation, make certain that the factory-prepared Configu-
ration file has been copied to the Default Data directory per instructions
on page 7 of manual. If your instrument has already been configured,
you can use the instructions in this section of the manual to check or, if
necessary, change the instrument settings.
1. Click Configure from the VISCPRO
®
menu bar.
2. Select your instrument group and instrument from the list of avail-
able instruments.
3. Select Tray Settings: Tube and Bath from the list of configuration
options. The Tray Settings: Tube and Bath window will appear.
The Tray Settings: Tube and Bath window
The Tray Settings: Tube and Bath window contains setup information
for each tube associated with your instrument. On multiple-tube CAN-
NON instruments, you can click on the tube tabs to see the setup infor-
mation for different tubes.
4. Verify that the tube serial number (Tube S/N) is correct. If it is not,
input the correct serial number in the text box.
Each tube should have
a unique serial number!
5. Verify the presence of calibra-
tion values (C and E) for each
bulb. If calibration data is not
available, the default values
are C=1 and E=0.
Sample calibration constants