When the Wash cycle has been completed, click Done to exit the Wash
viscometer tube window.
NOTES
The Wash operation cannot be executed if the tube is currently running a
sample. To terminate the Wash cycle at any time, click Abort Wash.
CAUTION
The Wash viscometer tube function overrides automatic miniAV-X
software operations. Verify that the tubes are clean and dry before
initiating computer-controlled sample testing.
To adjust the vacuum/pressure settings for the miniAV-X, click Service
from the VISCPRO menu options and select the desired instrument group
and instrument. Then click Adjust Vacuum/Pressure to open the Adjust
Vacuum/Pressure window.
Select the desired adjustment (vacuum or pressure) and then follow the
screen prompts to complete the adjustment procedure, using the appropri-
ate controls on the instrument wing panel. The prompts will display
recommended vacuum/pressure settings and will also indicate how
adjustments may be made to alter the settings.