Manual

CANNON
®
miniAV-X Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.Ob— March, 2012; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
68
The viscosity of the standard must be entered in the Check Standard
Viscosity field in order that the resultant viscosity determination may be
used for verification and calibration.
Copy & Paste Sample ID data entry options
The VISCPRO
®
Sample ID data entry screen supports Copy and Paste
options.
Copying Sample IDs
To copy a Sample ID entry, including the configuration information,
highlight the desired line(s) from the Sample List Box and right-click
your mouse to access the Sample Action window. Then select Copy All
Selected Samples from the menu options. OR you can use the C-C
key combination.
Pasting Sample IDs
To paste previously copied Sample ID entry information into another
tray/sample location within the same instrument group, highlight the
desired line(s) in the Sample List Box and right-click your mouse to
access the Sample Action window. Then select Paste Into Selected
Samples from the menu options. OR you can use the C-V key
combination.
Copying/pasting multiple IDs
You may copy or paste a block of (multiple) sample ID entries using the
same procedure described above. When pasting multiple entries, make
certain to select the same number of sample locations as were initially
copied. If you select a larger number, only the original copied number of
entries will be pasted. If you select a smaller number, information for the
last sample(s) in the block will not be copied.
Inserting/deleting a sample ID in the test sequence
To insert a sample ID into an existing sample run test sequence, highlight
the desired sample location in the sample run sequence and press I.
OR highlight the desired ID number, click the right mouse button, and
select Insert At from the Sample Action options.
The existing sample entries will be moved down one position in the
sequence and you can enter the new sample ID information at the cursor.
CAUTION
If the list box already contains a sample ID for every position on the tray,
the sample data for the final position will be discarded when a position is
cleared for new data.
To delete one or more sample IDs and move all remaining samples in the
sequence up in position, highlight the sample ID(s) you wish to delete
and press the D key. OR you may right-click in the highlighted block
and select Delete from the sample action options window. The high-