Manual

72
CANNON
®
miniAV
®
-X Automatic Viscometer with VISCPRO
®
Instruction & Operation Manual
Version 1.0— February, 2008; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
6. Click on the serial numbers in the Select Low Temperature Tubes
list box which correspond to tubes that should be configured as low
temperature tubes by the VI Matcher. As you do so, the serial num-
bers will be added to the Selected Tubes list box.
NOTE
Samples must have been previously run in a tube in order for the tubes
to show up in the Select Tubes list boxes
To “de-select” a selected tube, click again on the highlighted tube serial
number
7. Click on the tubes in the Select High Temperature Tubes list box
which correspond to tubes that should be configured as high tempera-
ture tubes by the VI Matcher. As you do so, the serial numbers will
be added to the Selected Tubes list box.
8. When you have finished adding low and high temperature tubes,
click OK to save the configuration and exit the General Configura-
tion window.
NOTES
To exit the General Configuration window without saving the configura-
tion, click Cancel.
For information on creating VI analyses, see the manual chapter corre-
sponding to the desired analysis.
Turning off the VI Matcher
The VI Matcher essentially function as a separate instrument and is
indicated as such on the VISCPRO
®
Status Bar. If you wish to terminate
VI matching, you will need to delete any VI domain(s) using the Delete
button from the General Configuration window. When no domains are
defined, the VI Matcher does not function.
Handling errors
VISCPRO
®
errors
The miniAV-X instrument software returns three types of errors during
operation: bath errors, fatal errors and tray errors. Bath and tray errors are
sample/tray specific, and may affect test results for a given sample. Such
results will be indicated as invalid. Fatal errors may be hardware-related,
and can produce a FAULT condition (see Chapter 5, Handling fault
conditions).