SimpleVIS Viscometer Reference Guide ® P82.
Run Status Indicator Lamps Sample/Solvent Port Dust Cover Waste Bottle Temperature Calibration Port Touch Screen Display RS-232 Serial Port Power Connector Exhaust Muffler Vacuum Line Waste Line
SAFETY CAUTIONS Please observe the following safety procedures and notices for proper operation of the SimpleVIS® Viscometer. • Make sure that you read and understand all operating instructions and safety precautions listed in this manual before installing or operating your unit. If you have questions regarding instrument operation or documentation, contact CANNON® Instrument Company. • Do not deviate from the installation, operation or maintenance procedures described in this guide.
~MAINS The ~MAINS symbol indicates instructions or connections for the AC power supply. The AC Power input must match the electrical specifications listed on the label on the rear panel of the instrument. The supplied AC Mains power cord must be attached to the connector labeled ~MAINS. This connection serves as a means of disconnect and should be readily accessible. The (O) symbol indicates the OFF position for the electrical switches for your unit (O) (AC Mains or accessories).
Running a Test GETTING STARTED Press the on switch on the AC/DC adapter. The Warm up Screen will display a “PLEASE WAIT” message. During this time, the instrument will seek the preset target temperature (40 or 100°C) and when reached, the instrument will enter an equilibration period. The equilibration period is 180 seconds. Note: Refer to the Basic Information section under Select Target Temperature for instructions on how to select the target temperature.
Running a Test Injecting a sample 1. Using the supplied piston operated pipettor, insert the tip into a vial of test sample and pull the measured amount into the pipettor. Be sure no bubbles exist in the oil or results may be erratic. Refer to pipettor instruction guide for more information. The pipettor volume setting should be set to 450 ~ 500 μL (0.45 - 0.50mL) 2.
Running a Test Print the results (only if optional printer is installed) 1. Press the Print button on the display screen to printout the test results to an attached serial printer (optional accessory). 2. You may skip the print option and go directly to the wash cycle. See the next section. Wash Cycle Result Screen 1.
Running a Test Dry Cycle Press the Dry button on the display wash screen to start the dry run cycle. The display will show the time remaining in the dry cycle. When the dry cycle time ends, the pump will turn off and the instrument will return to the Warm up Screen for a period of 60 seconds. The instrument will then be ready to run a new sample measurement at the leisure of the operator.
Maintenance & Calibration The SimpleVIS® Viscometer was designed to require minimal maintenance through simplicity of design and built-in electronic diagnostics. All new instruments are shipped with factory calibrations. However, like most metrological equipment, periodic maintenance and calibration is critical to maintaining accurate long-term measurements. CANNON® Instrument Company recommends the following intervals for maintaining the proper and accurate operation of the SimpleVIS® Viscometer.
Maintenance & Calibration Emptying the Waste Bottle Receiver Check the volume of the Waste Bottle frequently to avoid spillage/overflow. Disconnect the Waste Bottle from the cap containing the drain lines. Use appropriate precautions when handling the waste lines. Avoid spilling or splashing waste material collected in the lines. Check for kinks or wear in the lines. Always follow appropriate MSDS procedures when handling samples/solvents.
Maintenance & Calibration either of the two following calibration intervals. Performance Based Method Determine need to calibrate by performing weekly check-runs of certified viscosity reference standards. It is recommended the results of these runs be recorded and compared against the certified viscosity value printed on the standard bottle. When these check-runs consistently exceed the required precision (typically > 2 or 3%), then the instrument should be calibrated.
Maintenance & Calibration with the reference with the desired tolerance. Changing the temperature calibration by more than 0.05°C will affect the instrument viscosity calibration by 2% or more. Please consult CANNON® before proceeding. It is recommended that the instrument temperature be at target for at least 30 minutes before recalibrating if the instrument has been powered down for more than a couple of hours.
Basic Information Unpack your SimpleVIS® Viscometer Remove all packing materials from the components. The SimpleVIS® Viscometer should be placed on a stable laboratory table or bench. 1. Verify that you have received all components for the SimpleVIS® Viscometer. Report missing items to CANNON® Instrument Company immediately. 2. Inspect each component for signs of damage. Report damages to the shipper and CANNON® Instrument Company immediately. 3.
Basic Information vacuum line by approximately 2 inches. 4. Connect the other ends of both pieces of tubing as indicated in the photo image. Select Target Temperature 1. Press the on switch on the AC/DC adapter. The Warm up Screen will display a “PLEASE WAIT” message 2. Select the temperature by tapping the temperature directly under the “Seeking Target Temperature” caption field in the Warm Up Screen three times within two seconds. This will cause the display to switch to the Set Temperature Screen. 3.
Components for the SimpleVIS® Waste Bottle Viscometer Media Cap 9725-S10 P82.0075 Printer Cable Printer Power Printer with (optional) Supply Power Supply (optional) (optional) P82.0065 P82.0076 1/8” Transparent Tubing Plug Waste Bottle P65.3203 P81.2185 P82.0075 Pipettor P82.0055 Solvent Container P82.0053 AC/DC Adapter P82.0028 Pipette Seals, pkg. of 8 (optional) P82.0056 Power Supply Cord P74.2110 (120V) Pipette Tips (optional) P82.
2139 High Tech Road • State College • PA • 16803 814-353-8000 • 800-676-6232 • Fax 814-353-8007 cannon@cannoninstrument.com • www.cannoninstrument.com simplevismanual.indd 07.09.