Specifications

Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2-63
CALIBRATE:INTERNAL:STATUS?
might return :CALIBRATE:INTERNAL:STATUS UNCAL+ if non-volatile memory
has failed.
CALibrate:INTERNal:STARt
CH<x>? (Query Only)
Returns the vertical parameters of the specified channel. Because CH<x>:SCAle
and CH<x>:VOLts are identical, only CH<x>:SCAle is returned.
Vertical
CH<x>?
CH1?
might return the string :CH1:SCALE 10.0E-3;POSITION 0.0E0;
OFFSET 0.0E0;COUPLING DC;BANDWIDTH FULL;DESKEW 0.0E0;
IMPEDANCE MEG;PROBE 1.0E0;YUNIT ā€Vā€;INVERT 0 for channel 1.
CH<x>:BANdwidth
Sets or queries the bandwidth setting of the specified channel. This is equivalent
to setting Bandwidth in the Vertical menu.
Vertical
CH<x>:BANdwidth { TWEnty | ONEFifty | FULl }
CH<x>:BANdwidth?
<x> specifies the input channel. Valid values are 1, 2, 3, and 4.
TWEnty sets the channel bandwidth to 20 MHz.
ONEFifty sets the channel bandwidth to 150 MHz. This argument is not
available on some models.
FULl sets the channel bandwidth to the full bandwidth of the oscilloscope.
Examples
Related Commands
Group
Syntax
Examples
Group
Syntax
Arguments