TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2-69
<x> specifies the input channel. Valid values are 1, 2, 3, and 4.
<NR3> is the gain factor (output/input) of the probe. If the attached probe has
only one legal gain value gain, attempting to set the gain to another value results
in a Hardware Configuration error.
CH4:PROBE?
might return 1.0E-1 for a 10x probe.
CH<x>:SCAle
Sets or queries the vertical gain of the specified channel. This is equivalent to
setting Fine Scale in the Vertical menu or adjusting the front-panel Vertical
SCALE knob.
Vertical
CH<x>:SCAle <NR3>
CH<x>:SCAle?
<x> specifies the input channel. Valid values are 1, 2, 3, and 4.
<NR3> is the gain, in user units per division. The range is 10 V per division to
1 mV per division when using a 1x, 1 megohm, voltage probe.
CH4:SCALE 1.0E-1
sets the channel 4 gain to 10 units per division.
CH2:SCALE?
might return 1.0E+0, indicating that the current units per division setting of
channel 2 is 1 unit per division.