TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
DIAg:STATE EXECute (Set Only)
Executes the diagnostic tests that have been specified with the DIAg:SELect
command.
When the test sequence has completed, any of the modules or module interfaces
that failed diagnostics are displayed on the screen and stored in an internal log
file. The pass/fail status will be returned by the DIAg:RESUlt:FLAg? query, and
the internal log will be returned by the DIAg:RESUlt:LOG? query. This
command is equivalent to running Extended Diagnostics by selecting OK
Confirm Run Test when Execute is selected in the UTILITY menu and
System is set to Diags.
NOTE. The DIAg:STATE EXECute command can take 30 seconds or more to
respond. This command does not return control to the instrument controller until
diagnostics are complete.
Calibration and Diagnostic
DIAg:STATE EXECute
EXECute runs the diagnostic test sequences specified by the DIAg:SELect
command. When complete, the oscilloscope will return to the state it was in just
prior to the test. Pass/fail status of the tests can be returned by executing the
DIAg:RESUlt:FLAg? query.
DIAG:STATE EXECUTE
executes all the diagnostic tests that have been selected.
DISplay? (Query Only)
Returns the current display settings.
Display
DISplay?
DISPLAY?
returns to the current display settings.
might return :DISPLAY:FORMAT YT;STYLE VECTORS;FILTER SINX;PERSIST-
Group
Syntax
Arguments
Examples
Group
Syntax
Examples