TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 185
MASK:MASKPRE:WIDTH 648e-9
sets the User mask nominal bit width to 648 nanoseconds (1.544 Mbits/second).
MASK:MASKPRE:TRIGTOSAMP
MASK:POLarity
(Requires TDS3TMT)
Sets or returns the input waveform polarity for the pass/fail test. This command
only applies when MASK:TESt:STATE is on.
Mask
MASK:POLarity {BOTh | NEGAtive | POSITIVe }
MASK:POLarity?
BOTh sets the pass/fail test to test both positive and negative waveforms. The
oscilloscope tests all positive pulses until 1/2 of the waveforms have been tested.
Then the oscilloscope inverts all active channels and performs the remaining
tests on negative pulses.
NEGAtive inverts all active channels and performs the pass/fail test on negative
pulses.
POSITIVe performs the pass/fail test on positive pulses.
MASK:POLARITY BOTH
sets pass/fail to test both positive (normal) and negative (inverted) waveforms.
MASK:TESt:STATE
Examples
Related Commands
Group
Syntax
Arguments
Examples
Related Commands