Specifications

Command Descriptions
2- 186
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MASK:STANdard
(Requires TDS3TMT)
Set or returns the currently-displayed mask. To turn mask display off, set the
standard to NONe.
Mask
MASK:STANdard { CLOCKCoax | CLOCKSymmetrical | DS0Contra
| DS0Double | DS0Single | DS0Timing | DS1C | DS1Rate | DS2
| DS2RATECoax | DS2RATESymmetrical | DS3Rate | E1Coax
| E1Symmetrical | E2 | E3 | G703DS1 | G703DS3 | NONe | RATE32Mbit
| STS1Pulse| USERMask }
MASK:STANdard?
CLOCKCoax through USERMask are the available TDS3TMT masks.
MASK:STANDARD STS1PULSE
sets the current mask to STS1Pulse.
MASK:STOPOnviolation
(Requires TDS3TMT)
Sets or returns the stop on violation setting. When enabled, this command stops
waveform acquisitions on the first occurance of a waveform violation.
MASK:DISplay must be ON for the oscilloscope to detect violations.
Mask
MASK:STOPOnviolation { OFF | ON | <NR1> }
MASK:STOPOnviolation?
OFF turns off mask test stop on violation.
ON stops waveform acquisition on the first occurance of a mask violation.
<NR1> is an integer number where NR1 = 0 turns off stop on violation, and
NR1 0 turns on stop on violation.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments