Specifications

Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 211
MEASUrement:IMMed:TYPe
Specifies the immediate measurement.
Measurement
MEASUrement:IMMed:TYPe { AMPlitude | AREa | BURst | CARea |
CMEan | CRMs | DELAY | FALL | FREQuency | HIGH | LOW | MAXimum |
MEAN | MINImum | NDUty | NOVershoot | NWIdth | PDUty | PERIod |
PHASE | PK2pk | POVershoot | PWIdth | RISe | RMS }
MEASUrement:IMMed:TYPe?
Information on how measurements are derived:.
AMPlitude is the high value minus the low value.
AREa is the area between the curve and ground over the active waveform the high
value minus the low value. TDS3AAM only.
BURst is the time from the first MidRef crossing to the last MidRef crossing.
CARea (cycle area) is the area between the curve and ground over one cycle.
TDS3AAM only.
CMEan is the arithmetic mean over one cycle.
CRMs is the true Root Mean Square voltage over one cycle.
DELay is the delay from one waveform’s edge event to another. TDS3AAM only.
FALL is the time that it takes for the falling edge of a pulse to fall from a HighRef
value to a LowRef value of its final value.
FREQuency is the reciprocal of the period measured in hertz.
HIGH is the 100% reference level.
LOW is the 0% reference level.
MAXimum is the highest amplitude (voltage).
MEAN for general purpose measurements, is the arithmetic mean over the entire
waveform. For histogram measurements, it is the average of all acquired points
within or on the histogram box.
MINImum is the lowest amplitude (voltage).
NDUty is the ratio of the negative pulse width to the signal period expressed as a
percentage.
Group
Syntax
Arguments