TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MEASUrement:INDICators:VERT<x>? (Query Only)
Returns the value of the specified vertical measurement indicator <x> from the
trigger point, where <x> can be 1, 2, 3, or 4. A negative value means that the
indicator is positioned earlier in the waveform record than the trigger point.
Measurement
MEASUrement:INDICators:VERT<x>?
MEASUREMENT:INDICATORS:VERT2? might return
:MEASUREMENT:INDICATORS:VERT2 -3.724507E-6 indicating that the second
measurement indicator is positioned 3.72 ī€ƒs before the trigger point.
MEASUrement:MEAS<x>? (Query Only)
Returns all measurement parameters for the specified active measurement <x>.
Measurement
MEASUrement:MEAS<x>?
Measurement parameters, for the specified measurement.
MEASUREMENT:MEAS3?
might return: :MEASUREMENT:MEAS3:TYPE PERIOD;UNITS ā€sā€;COUNT
0;MINIMUM 0.0E0;MAXIMUM 0.0E0;MEAN 0.0E0;STDDEV 0.0E0;SOURCE1
CH1;SOURCE2 CH2;DELAY:DIRECTION FORWARDS;EDGE1 RISE;:MEASURE-
MENT:MEAS3:STATE 0
MEASUrement:MEAS<x>:COUNt? (Query Only)
(Requires TDSAAM)
Returns the number of values accumulated for this measurement since the last
statistical reset. Some values may have been ignored because they generated an
error.
Measurement
Group
Syntax
Examples
Group
Syntax
Returns
Examples