TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MEASUrement:MEAS<x>:VALue? (Query Only)
Returns the value that has been calculated for the measurement specified by <x>.
NOTE. This value is a display value and is updated on the oscilloscope screen
approximately every 1/3 second.
Measurement
MEASUrement:MEAS<x>:VALue?
MEASUREMENT:MEAS1:VALUE?
might return MEASUREMENT:MEAS1:VALUE 3.742E-6, indicating the measure-
ment value displayed at the time the command was executed.
MEASUrement:MEAS<x>:DATa?
MEASUrement:METHod
Sets or returns the method used to calculate the 0% and 100% reference level.
This is equivalent to setting the High-Low Setup in the MEASURE menu.
Measurement
MEASUrement:METHod { Auto | HIStogram | MINMax }
MEASUrement:METHod?
HIStogram sets the high and low waveform levels statistically using a histogram
algorithm.
MINMax sets the high and low waveform levels to MAX and MIN, respectively.
Auto chooses the best method (HIStogram or MINMax) for each dataset.
Group
Syntax
Examples
Related Commands
Group
Syntax
Arguments