TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
MEASUREMENT:REFLEVEL:METHOD ABSOLUTE
specifies that explicit user-defined values are used for the reference levels.
MEASUREMENT:REFLEVEL:METHOD?
returns either ABSolute or PERCENT, indicating the reference levels used.
MEASUrement:REFLevel:PERCent:HIGH
Sets or returns the percent, where 100% is equal to HIGH, that is used to
calculate the high reference level when MEASUrement:REFLevel:METHod is
set to PERCent. This command is equivalent to setting the relevant Reference
Levels in the MEASURE menu.
Measurement
MEASUrement:REFLevel:PERCent:HIGH <NR3>
MEASUrement:REFLevel:PERCent:HIGH?
<NR3> ranges from 0 to 100 percent, and is the high reference level.
The default is 90%.
MEASUREMENT:REFLEVEL:PERCENT:HIGH 95
specifies that the high reference level is set to 95% of HIGH.
MEASUrement:REFLevel:PERCent:LOW
Sets or returns the percent, where 100% is equal to HIGH, that is used to
calculate the low reference level when MEASUrement:REFLevel:METHod is set
to PERCent. This command is equivalent to setting the relevant Reference
Levels in the MEASURE menu.
Measurement
MEASUrement:REFLevel:PERCent:LOW <NR3>
MEASUrement:REFLevel:PERCent:LOW?
<NR3> ranges from 0 to 100 percent, and is the low reference level.
The default is 10%.
Examples
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments