Specifications

Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 271
TRIGger:A:HOLdoff:VALue
Sets or queries the A trigger holdoff time as a percent of record. This is
equivalent to setting Holdoff (% of record) in the Mode & Holdoff side menu.
Trigger
TRIGger:A:HOLdoff:VALue <NR3>
TRIGger:A:HOLdoff:VALue?
<NR3> from 0 to 100 and is a percent of the holdoff range.
TRIGGER:A:HOLDOFF:VALUE 50
to set the holdoff value to 50% of record.
TRIGger:A:LEVel
Sets the A trigger level. This command is equivalent to adjusting the front-panel
Trigger LEVEL knob.
Trigger
TRIGger:A:LEVel { ECL | TTL | <NR3> }
TRIGger:A:LEVel?
ECL specifies a preset ECL level of --1.3 V.
TTL specifies a preset TTL level of 1.4 V.
<NR3> is the A trigger level, in volts.
TRIGGER:A:LEVEL?
might return :TRIGGER:A:LEVEL 1.4E0, indicating that the A edge trigger level
is set to 1.4 V.
TRIGGER:A:LEVEL 0.5
sets the A trigger level to 0.5 V.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments
Examples