Specifications

Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 277
TRIGGER:A:LOGIC:PATTERN:WHEN LESSTHAN
sets the oscilloscope to trigger when the pattern is true for a time period less than
the pattern trigger delta time setting.
TRIGger:A:LOGIc:STATE:WHEn
Sets or returns the state condition on which to trigger the oscilloscope.
Trigger
TRIGger:A:LOGIc:STATE:WHEn { TRUe | FALSe }
TRIGger:A:LOGIc:STATE:WHEn?
TRUe triggers the oscilloscope if the state signal is true when the clock signal
slope is true.
FALSe triggers the oscilloscope if the state signal is false when the clock signal
slope is true.
TRIGGER:A:LOGIC:STATE:WHEN?
might return TRIGGER:A:LOGIC:STATE:WHEN TRUE.
TRIGger:A:MODe
Sets or queries the A trigger mode. This command is equivalent to selecting
Mode & Holdoff in the Trigger menu.
Trigger
TRIGger:A:MODe { AUTO | NORMal }
TRIGger:A:MODe?
AUTO generates a trigger if a trigger is not detected within a specific time period.
NORMal waits for a valid trigger event.
TRIGGER:A:MODE AUTO
specifies that a trigger event is automatically generated.
Examples
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments
Examples