TRIGger:A:PULse:SLEWRate:THReshold:HIGH
Sets or returns the slew rate source signal high threshold voltage value.
Trigger
TRIGger:A:PULse:SLEWRate:THReshold:HIGH <NR3>
TRIGger:A:PULse:SLEWRate:THReshold:HIGH?
<NR3> is a floating point value with exponent that sets the slew rate trigger signal
high threshold voltage value.
TRIGGER:A:PULSE:SLEWRATE:THRESHOLD:HIGH 4.1E0
sets the slew rate high threshold voltage value to 4.1 volts.
TRIGger:A:PULse:SLEWRate:THReshold:LOW
Sets or returns the slew rate source signal low threshold voltage value.
Trigger
TRIGger:A:PULse:SLEWRate:THReshold:LOW <NR3>
TRIGger:A:PULse:SLEWRate:THReshold:LOW?
<NR3> is a floating point value with exponent that sets the slew rate trigger signal
low threshold voltage value.
TRIGGER:A:PULSE:SLEWRATE:THRESHOLD:LOW -1.5E0
sets the slew rate low threshold voltage value to --1.5 volts.