Specifications

Command Descriptions
2- 286
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
TRIGger:A:PULse:SLEWRate:WHEn
Sets or returns the slew rate condition on which to trigger the oscilloscope.The
condition is based on the slew rate calculated from the values set by the
TRIGger:A:PULse:SLEWRate:DELTATime and TRIGger:A:PULse:SLEW-
Rate:THReshold commands.
Trigger
TRIGger:A:PULse:SLEWRate:WHEn { FASTERThan | SLOWERthan
| EQual | NOTEQual }
TRIGger:A:PULse:SLEWRate:WHEn?
FASTERThan triggers the oscilloscope when the signal slew rate is faster than the
calculated slew rate.
SLOWERthan triggers the oscilloscope when the signal slew rate is slower than
the calculated slew rate.
EQual triggers the oscilloscope when the signal slew rate is equal to the
calculated slew rate within a ±5% tolerance.
NOTEQual triggers the oscilloscope when the signal slew rate is not equal to the
calculated slew rate within a ±5% tolerance.
TRIGGER:A:PULSE:SLEWRATE:WHEN NOTEQUAL
sets the oscilloscope to trigger when the slew rate is not equal to the calculated
slew rate.
TRIGger:A:PULse:SOUrce
Sets or returns the signal source for width, runt, and slew rate pulse triggers. This
command is equivalent to selecting a trigger signal source in the Source side
menu.
Trigger
Group
Syntax
Arguments
Examples
Group