Specifications

Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 299
TRIGger:B:EDGe:SLOpe
Selects a rising or falling slope for the B edge trigger. This is equivalent to
setting Slope in the Trigger menu.
Trigger
TRIGger:B:EDGe:SLOpe { FALL | RISe }
TRIGger:B:EDGe:SLOpe?
FALL specifies to trigger on the falling or negative edge of a signal.
RISE specifies to trigger on the rising or positive edge of a signal.
TRIGGER:B:EDGE:SLOPE RISE
sets the B edge trigger to occur on the rising slope.
TRIGger:B:EDGe:SOUrce
Sets or queries the source for the B edge trigger. This is equivalent to setting
Source in the Trigger menu.
Trigger
TRIGger:B:EDGe:SOUrce { LINE | CH<x> | EXT | EXT10 }
TRIGger:B:EDGe:SOUrce?
LINE sets the B trigger source to the AC line frequency. LINE is not available
when the instrument is operating on battery power.
CH<x> sets the B trigger source to the specified input channel, where <x> is 1, 2,
3, or 4.
EXT sets the B trigger source to the regular external trigger input connector with
a signal input range of --0.8 V to +0.8 V. EXT is not available in 4-channel
TDS3000 Series instruments.
EXT10 sets the B trigger source to the reduced external trigger with a signal input
range of --8 V to +8 V. EXT10 is not available in 4-channel TDS3000 Series
instruments.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments