Specifications

Command Descriptions
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
2- 321
WFMPre:XINcr <NR3>
WFMPre:XINcr?
<NR3> is the sampling interval in seconds
or Hertz per point.
WFMPre:<wfm>:XINcr
WFMPre:XUNit
The set form of this command is only valid for reference waveforms. The only
valid units are “s” and “Hz”. If an attempt is made to set the XUNit to something
other than “s” or “Hz”, event 224 (illegal parameter value) is generated. Setting a
reference waveform’s XUNit to Hz causes the oscilloscope to display the
reference waveform as an FFT waveform.
The query form returns the horizontal units for the waveform specified by
DATa:SOUR ce (“s” for seconds, “Hz” for Hertz). If the waveform specified by
DATa:SOUR ce is not displayed, the query fails and the oscilloscope generates an
execution error with event code 2244 (waveform requested is not on).
Waveform
WFMPre:XUNit <qstring>
WFMPre:XUNit?
<qstring> is “s” or “Hz”.
WFMPre:<wfm>:XUNit
Syntax
Arguments
Related Commands
Group
Syntax
Arguments
Related Commands