TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
ALLEv? (Query Only)
Causes the oscilloscope to return all events and their messages, and removes the
returned events from the Event Queue. The messages are separated by commas.
Use the *ESR? query to enable the events to be returned. For a complete
discussion of the use of these registers, see page 3--1. This command is similar to
repeatedly sending *EVMsg? queries to the instrument.
Status and error
ALLEv?
The event code and message in the following format:
<Event Code><Comma><QString>[<Comma><Event Code><Comma>
<QString>...]
<QString>::= <Message>;[<Command>]
<Command> is the command that caused the error and may be returned when a
command error is detected by the oscilloscope. As much of the command will be
returned as possible without exceeding the 60 character limit of the <Message>
and <Command> strings combined. The command string is right-justified.
ALLEV?
might return the string :ALLEV 2225,”Measurement error, No waveform to
measure; ”,420,”Query UNTERMINATED; .
*CLS, DESE, *ESE, *ES R?, EVENT?, EVMsg?, EVQty, *SRE, *STB?
AUTOSet (Set Only)
Sets the oscilloscope to automatically adjust its vertical, horizontal, and trigger
controls to provide a stable display of the selected waveform. This is equivalent
to pressing the front-panel AUTOSET button.
Acquire
AUTOSet { EXECute | UNDO | VIDeo | VFields | VLines }
Group
Syntax
Returns
Examples
Related Commands
Group
Syntax