Specifications

Command Descriptions
2-60
TDS3000, TDS3000B, and TDS3000C Series Programmer Manual
CALIBRATE:FACTORY:NOTIFY:YEARS 1.5e0
sets the calibration due time to 1.5 years.
CALIBRATE:FACTORY:NOTIFY:YEARS?
might return :CALIBRATE:FACTORY:NOTIFY:YEARS 1.5E0 indicating that the
calibration due notifier will be displayed 1.5 years after the last calibration.
CALibrate:FACtory:PREVious (Set Only)
This command backs up the calibration steps one step so that the next CALi-
brate:FACtory:CONTInue command will repeat the factory calibration step that
just finished. Repeated CALibrate:FACtory:PREVious commands can back up as
many calibration steps as desired (until step one is reached). The appropriate
notifier (with signal source information) is displayed after the factory calibration
step is adjusted.
Calibration and Diagnostic
CALibrate:FACtory:PREVious
CALIBRATE:FACTORY:PREVIOUS
backs up the calibration steps one step.
CALibrate:FACtory:STARt (Set Only)
The factory calibration process consists of a series of steps. This command
initializes the factory calibration sequence, but performs no actual signal path
characterization.
The calibration process automatically sets up the oscilloscope for each step. Do
not send any other commands that change the state of the oscilloscope until the
calibration process is complete. Send the CALibrate:FACtory:CONTInue
command to begin the selected calibration step (after the appropriate signal
source is connected). Send CALibrate:ABOrt or *RST to abort the calibration
process.
Calibration and Diagnostic
CALibrate:FACtory:STARt
Examples
Group
Syntax
Examples
Group
Syntax