Specifications

– 14 –
4.2 Measurement in FULL AUTO Mode
4
Start automatic measurement
Press the START switch.
Measurement head moves automati-
cally to perform alignment.
When a warning message “[1] EYE CANNOT BE FOUND PRESS START
SW AGAIN” is displayed, or when alignment does not end
When automatic alignment cannot be performed within a specific time, the measure-
ment head stops moving and a warning message “[1] EYE CANNOT BE FOUND
PRESS START SW AGAIN” will be displayed. Also, in some cases, alignment does
not end depending on the condition of examinee’s eye.
In this case, take the measures indicated below according to the cause.
[Cause 1] The position of examinee’s eye is off the automatic alignment range.
[Remedy 1] Move the position of the examinee’s eye displayed on the monitor toward
the center a little more using the roller and trackball. Then, press the
START switch. Adjust the height of the examinee’s eye with the height of
the chin rest as required.
[Cause 2] Examinee’s eye is moving restlessly due to nystagmus, etc.
[Remedy 2] Have him/her look at the eye fixation target and press the START switch
again.
If the eye is moving due to causes such as nystagmus, perform compul-
sory measurement by pressing the START switch during automatic
alignment (see below), or perform continuous manual measurement by
pressing the START switch for about 2 seconds ( next page).
– Compulsory measurement –
In compulsory measurement, each time the START switch is pressed,
measurement is performed once.
NOTE: Measurement error will occur unless the positioning is correct.
PD: – – VD:13.5 No. 00001
FULL AUTO(3)
RIGHT SPH CYL AX
1:
[ ]
Corneal bright dots