Operating Instructions Temperature Indicator Eurotherm 2132 or 2116 CONTENTS section 0 2 2 3 page Description Operation User Calibration Navigation Diagram 2 2 3 4 MC06 – 1.
132/2116 Indicator Description The model 2132 or 2116 controller/indicator is made by Eurotherm, and is fitted by Carbolite configured for immediate use as an indicator. It is configured for the input (thermocouple) type for which it was ordered. The indicator does not contain a real-time calendar, and is not subject to century-end date problems. When the indicator is supplied fitted into a furnace or oven then it is usually on whenever the Instrument Switch is on.
2132/2116 Indicator 2 USER CALIBRATION The indicator is calibrated for life at manufacture against known reference sources, but there may be sensor errors or other system errors. User calibration allows compensation for such errors, and the 2132 or 2116 allows for a user 2-point calibration. This setting is password protected to avoid accidental alteration. Page to iP, scroll to CAL.P, and use Up to alter the password. The password is 3. If the correct password is entered, the display shows PASS.
3 NAVIGATION DIAGRAM Home List 20.0 Input List *C iP Access List ACCS *C CAL.P codE enter password measured temperature for factory access to lists and parameters not available to the operator CAL if user calibration Pnt.L user 2-point calibration OFS.L Pnt.H OFS.H For preventive maintenance, repair and calibration of all Furnace and Oven products, please contact: Carbolite Engineering Services Telephone: +44 (0)1433 624242 Fax: +44 (0)1433 624243 Email: service@carbolite.