User guide

PTC
MF11 3.08 15
8.0 CIRCUIT DIAGRAM
8.1 Circuit Diagram
9.0 FUSES & POWER CONTROL
9.1 Fuses
Supply Fuses: 32mm x 6mm type F1 (quick blow)10 Amp. Access by removal of the furnace back
panel.
Protection Fuse: 20mm x 5mm type FF (ultra rapid) 10 Amp. Access by removing the holder
underneath the furnace, using a screwdriver.
The purpose of the Protection Fuse is to ensure immediate power cut off if the voltage change
switch is in the 115V position and the furnace is connected to a 220-240V supply.
9.2 Power Control
In this model the same heating element is used for 110-120V models and 220-240V models.
Changing between the voltages is by the switch underneath the furnace see section 2.2.
The correct power limit setting (parameter OP.Hi) for the furnace for all voltages is 100%.
Instrument Switch
temperature
controller
thermocouple
SSR
e
l
e
m
e
n
t
heat
on
Supply
Fuses
N
L
PE
temperature
indicator
thermocouple
e
l
e
m
e
n
t
voltage
change
switch
(shown in
230V
position)
Protection
Fuse