8.2 Power Settings
The power limit settings (parameter OP.Hi) for the RHF models on various supply voltages are as
follows. The figures represent the maximum percentage of time that controlled power is supplied
to the elements. Do not attempt to “improve performance” by setting a value higher than the
correct one from the table.
Model Volts:
200V
208V
220V
230V
240V
380V
400V
415V
RHF 14/3
90
83
74
68
63
74
68
63
RHF 15/3
62
57
51
47
43
51
47
43
RHF 16/3
62
57
51
47
43
51
47
43
RHF 14/8
90
83
74
68
63
74
68
63
RHF 15/8
60*
56*
76
69
63
76
69
63
RHF 16/8
73*
68*
74
68
62
74
68
62
RHF 14/15
92
85
76
69
64
76
69
64
RHF 15/15
71
66
59
54
49
59
54
49
RHF 16/15
73
68
61
55
51
61
55
51
RHF 14/35
93
86
77
71
65
77
71
65
RHF 15/35
93
86
77
71
65
77
71
65
RHF 16/35 (star)
-
-
-
-
-
85
78
71
RHF 16/35 (delta)
-
89
79
73
67
-
-
-
* these models use different elements in 200-208V versions