User's Manual

Page10of17
TECHNICAL SPECIFICATIONS
The technical specications are subject to change.
PHYSI CAL
Sensor 3 inches x 1.9 inches x 0.7 inch; Weight: 3.0 oz. with battery
Monitor 4.7 inches x 2.6 inches x 0.9 inch; Weight: 6 oz.
LCD 2.27 inches x 1.7 inches; Touch screen: color
Base 4.3 inches x 3.7 inches x 1.0 inches; Weight: 6.0 oz.
FUNCTI ONAL
Sample Rate 250 samples per second
Resolution 12 bits
Dynamic range + / - 5 mV
Bandwidth 0.1 to 40 Hz
Channels 2
Battery Life: Monitor Minimum 10 hrs (with cleared memory & fully recharged battery)
Battery Life: Sensor 24 hrs (1 AAA Alkaline)
Leakage Current Less than .1 µ A Electrodes
TRANSMI SSI ON
Sensor to Monitor 900 MHz I SM band RF transmission, digital error corrected. 150 foot range.
Retransmission if data is corrupted.
Monitor to Center CDMA (PCS and cellular) wireless, digital error corrected. Telephone line
modem, digital error corrected. CDMA (PCS & cellular) wireless, digital error
corrected. Telephone line modem, digital error corrected. ECG recording
intervals vary by type of event; the minimum is 15 seconds pre-event and
45 seconds post event. Wireless transmission subject to coverage of cellular
network. Data is received at Monitoring Center within minutes with good
wireless coverage and/ or modem communications.
OPERATI NG CONDI TI ONS
Operating Temperature
Sensor: 20 - 45°C; Monitor ; 0 - 45°C
Operating Humidity 10% - 95% non-condensing
Storage Temperature
-20 - 65°C non-condensing
Storage Humidity 5% - 95% non-condensing
Operation Altitude 700 - 1060 millibars