Calibration Adjustment Mode
Bio-medical technicians. It will alter the factory
calibration of Caregiver®.
The displayed temperature in SURFACE mode may be increased or
decreased by an amount up to ±2°F maximum in increments of
0.1°F. This will also alter the displayed reading in BODY mode but
not necessarily by the same amount.
Details on the recommended calibration interval, checking
calibration and adjusting calibration are found at
www.thermomedics.com in the knowledge base under calibration.
Displayed
Temperature Range
BODYMode:94˚Fto108.0˚F(34.4˚Cto42.2˚C)
SURFACEMode:32˚Fto140˚F(0˚Cto60˚C)
Accuracynotspecifedbelow71.6˚F(22˚C)
Storage
Temperature Range
-4˚Fto140˚F(-20˚Cto+60˚C)
Ambient Operating
Range
50˚Fto104˚F(10˚Cto40˚C)
85% Relative Humidity or less.
Displayed Accuracy From96.8/102.2˚F±0.4˚F(36to39˚C±0.2˚C)
From71.6/96.7˚F±0.5˚F(22to35.9˚C±0.3˚C)
From102.3/108.0˚F±O.5˚F(39.1to42.2˚C±0.3˚C)
Power Two 1.5v “AA” Alkaline Batteries
Battery Life > 10,000 Temperatures Typical
Response Time Approx 1 second
Size (LxWxH)150x48.48x55.16mm
Batteries 3V 2XAA
Type BF Equipment
This device has been tested and conforms to
the standard ASTM E1965-98:2009.