Product Info

CS 1500 Intraoral Camera Safety and Regulatory Guide
(9H0953)_Ed06 5
Hygiene and Disinfection
Perform the following maintenance activities on the CS 1500 and accessories
regularly.
To ensure maximum hygienic safety for the patient, carefully follow the
instructions to prepare the CS 1500 camera for use.
To minimize the risk of cross-contamination, after each patient, clean and
disinfect the CS 1500. See “Cleaning and Disinfecting the CS 1500” on
page 6.
General Warnings
WARNINGS:
Read and follow the warnings and personal
protection instructions provided in the Safety
Data Sheet (SDS) of the disinfectant before
use.
The disinfectant should be approved by the
applicable competent authority for use on a
dental medical device.
You must wear gloves while handling and
using the camera.
Always protect the camera with a hygienic
barrier sheath before use.
The camera should be disinfected with a U.S.
Environmental Protection Agency (EPA)-
registered or CE-marked intermediate-level
disinfectant solution with tuberculocidal
activity between patients.
DO NOT use a disinfectant containing
phenolics or iodophors; doing so will damage
the surface coating of the camera.
Never put the camera in a sterilizing device or
immerse it in water or the disinfectant
solution.
Excessive fluids can damage the camera.
Not protected against water spray.