84 Chapter 11 Troubleshooting
5 Unscrew the Phillips screw that attaches the service door to the rear cover, using the Phillips PH2
screwdriver and remove the service door.
6 Tilt the scanner backwards, until the plate falls out of the scanner into your hand.
7 After the imaging plate has been removed, replace the rear service cover and reinstall the crosshead
screw firmly so that there is no possibility of light penetration.
8 Connect the scanner to the mains power supply.
9 Reconnect the scanner to the network.
10 Turn on the scanner.