User Manual

8 9G4259_en 2012-02-24
Safety and Related Information
Canada
Europe and the Rest of World
Under 60601-1-2, there are four tests for 61000-4-11. Three of the tests
were passed normally at the declared test voltages.
Compliance with one of the tests, a 60 % dip in voltage for 5 cycles
(which is approximately one tenth of a second) was not met due to a
technological limitation. The imager portion of the CHROMA Imaging
System will not function at the minimum nominal voltage of 100 V,
when it dips to 40 V. At 40 V, the imager stops and requires operator
intervention to be restored to full operation (Test Performance Criteria
“C”. However, the imager must meet Test Performance Criteria “B” or
“A”). Although the imager stops and intervention is required, no images
are lost by the imaging system.
The imager passes the 60 % dip test at 115 V nominal, dipping to 46 V.
No operator intervention is required, and there are no performance
degradation results due to the voltage dip (Test Performance
Criteria “A”).
The DICOM interface passes the 60 % dip test at 100 V nominal,
dipping to 40 V (Test Performance Criteria “A”). Therefore, the
DICOM interface may be plugged directly into a 100 V system, wall
outlet, and meets all the requirements of 61000-4-11 under 60601-1-2.
CAN/CSA-C108.6-M91, Class A: Limits and Methods of
Measurement of Electromagnetic Disturbance Characteristics of
Industrial, Scientific and Medical (ISM) Radio-Frequency
Equipment.
Intentional Radiation IC for CHROMA Imager: 7027A-4900.
This Class A digital apparatus complies with Canadian ICES-003.
CET APPAREIL NUM ENRIQUE DE CLASSE A EST
CONFORME A LA NORME NMB-003 DU CANADA.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
SHORT TERM CONFIDENTIAL